400 User Manual
Combustion Test
NGAS
Acme Powerplant
Fig. 3-7:
Combustion Test Screen
Fuel Type -
displays fuel calculation being used to
produce combustion data.
Pump Status -
( ) indicates that the pump is on,
( ) indicates that the pump is off.
Folder / Memory Location -
location where
combustion data will be saved.
Combustion Data -
results from combustion
analysis; display can be toggled by pressing ( ).
Run / Hold -
toggle sample pump operation; press ( )
to run and ( ) to pause.
Toggle View -
change data display between table /
list ( ), line graph ( ), tile ( ) formats.
Options -
displays additional settings / configuration
(Fresh Air Purge, Logging, Pump Speed, Run / Hold Format,
Line Graph Settings, O
2
Reference, Zoom, NO
x
Calculation)
.
Save -
save combustion data to the selected folder /
memory location.
Print -
print combustion data to IrDA / Bluetooth
®
printer(s).
IMPORTANT
:
Printing via Bluetooth
®
is diabled when the analyzer is connected to
the PCA
®
400 App. If you wish to print, disconnect from the mobile app or enable
IrDA communications.
5. Run Sample Pump
After starting the pump, allow readings to stabilize for a minimum of 3-minutes or as long
as required by manufacturer specifications / compliance testing procedures.
f
To start the pump: press
Run / Hold
( )
f
To stop the pump: press
Run / Hold
( )
6. Save / Print Readings
The PCA
®
400 is capable of saving active data to the analyzer’s onboard storage and / or
printing to the optional IrDA / Bluetooth
®
printer.
f
To save combustion readings / report: press Save ( )
f
To print combustion readings / report: press Print ( )