Bacharach ODOROMETER Manual Download Page 6

Page-6 

 Instruction  0023-9125

ODOROMETER

The instrument has two ranges, obtained by using a lightweight glass 

and a heavier metal 

fl

 oat, both in the same metering tube. Take readings 

on the glass 

fl

 oat when both are within the range of measurement; take 

readings on the metal 

fl

 oat when the glass 

fl

 oat reaches the top of the tube. 

Useful ranges are approximately 0.04 to 0.4% for the glass 

fl

 oat, and 0.2 

to 1.1% for the metal 

fl

 oat.

When taking readings, observe the FLOWMETER graduation corresponding 

to the bottom of the ball 

fl

 oat. Find the gas concentration corresponding to 

this reading by looking at the calibration chart on the inside of the front 

cover. This chart has been drawn to give the percent gas concentration over 

the useful ranges for natural gas with a speci

fi

 c gravity of 0.620 relative 

to air.  Concentrations of  gases of other  densities may  be read using  the 

correction factors shown on the graph below the calibration chart.

The 

FLOWMETER

 tube and 

fl

 oats must be kept clean to prevent 

sticking of the 

fl

 oats in the tube. To disassemble, press the sides of the 

plastic cover and pull out. Remove the spring lock clip at the top of tube, 

lift the tube connector and pull tube out, by lifting from the bottom con-

nectors. Tube and 

fl

 oats can be washed in alcohol, trichloroethylene or 

other solvent.

When the odor (mercaptan) in gas has been detected, observe the reading 

indicated by the FLOWMETER 

fl

 oat. Depending on the range required, 

use either the upper glass 

fl

 oat (colored) for values 0.04 to 0.4% or lower 

metal 

fl

 oat (silver) for values 0.2 to 1.2%.

EXAMPLE: FLOWMETER (silver 

fl

 oat) indicates 7; use second curve and 

observe that percent gas indicated by the chart is 1.1% (% 

GAS IN ODOROMETER EFFLUENT).  If speci

fi

 c gravity of 

gas under test  is 0.620, the correct percentage of gas is as 

noted on the curve (1.1%). If speci

fi

 c gravity differs, use the 

CORRECTION CHART below.

EXAMPLE: Speci

fi

 c gravity of gas under test is 0.55, the correction factor 

noted on the chart is 1.06. The correct gas concentration is 

obtained by multiplying the following:

          1.1% x 1.06 = 1.166% (Gas in ODOROMETER Ef

fl

 uent).

                         

6. 

RANGE OF  

 

 MEASUREMENT

7. MAINTENANCE

8. 

HOW TO USE THE  

 

ODOROMETER GAS  

 CALIBRATION 

 

  

CHART

Summary of Contents for ODOROMETER

Page 1: ...information in this manual must be fully understood and utilized to ensure that the instrument is operating prop erly and is both used and maintained in the proper manner by qualified personnel An instrument that is not properly calibrated operated and maintained by qualified personnel is likely to provide erroneous infor mation which could prevent user awareness of a potentially hazardous situati...

Page 2: ... Bacharach to the replacement of this Product or parts thereof that are defective in title The warranty set forth in paragraph 1 does not apply to parts the Operating Instructions designate as having a limited shelf life or as being expended in normal use THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF i ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT...

Page 3: ...edle valve The concentrationofgasintheeffluentairisdeterminedfromtheFLOWMETER reading by reference to the calibration chart on the inside of the front cover The instrument components are mounted on an anodized aluminum panel which in turn mounts inside an anodized aluminum case All metal and instrument surfaces have been chosen so that no outgas problems will occur Make sure not to use any chemica...

Page 4: ...tests should be run in an odor and draft free area 4 Set the instrument on a level bench or desk Swing the handle down and the top cover open Switch the unit ON Clockwise the green POWER LED should light See Appendix A 2 for older models NOTE If the LED goes out after a few seconds replace the batteries 5 The GAS INLET NEEDLE VALVE is located in the center of the panel Open the valve slowly while ...

Page 5: ... the threshold and to run an occasional blank using air with no added gas 5 Calculate the average threshold value of each observer Then calculate the total average value using these individual averages This procedure is advisable because considerable variation is found among individuals in detecting threshold value Thus if the observ ers take different numbers of readings a straight average of all...

Page 6: ...n to prevent sticking of the floats in the tube To disassemble press the sides of the plastic cover and pull out Remove the spring lock clip at the top of tube lift the tube connector and pull tube out by lifting from the bottom con nectors Tube and floats can be washed in alcohol trichloroethylene or other solvent When the odor mercaptan in gas has been detected observe the reading indicated by t...

Page 7: ...ODOROMETER GAS CALIBRATION CHART This chart is intended as an example only and does not reflect the cali bration of your instrument See actual calibration graph on your instru ment Your ODOROMETER was calibrated against standards traceable to ...

Page 8: ...erval will be based upon the instrument s condition as received in or out of tolerance and degree of drift since the last calibra tion Bacharach recommends that service on portable instruments in or out of warranty be performed at one of our Service Centers Should service be required please pack the instrument carefully include all necessary identifying papers with description of service needed an...

Page 9: ...any 115 volt 60 Hz power source 2 For the DC model start motor by turning switch clockwise to ON and rotating the knob clockwise until air flow is achieved Set am meter to red mark for proper amount of air flow For the AC model just plug in line cord and turn on switch to start air flow Pilot light will indicate that power is connected to motor ...

Reviews: