Bacharach Neutronics LEGEND ID RI-2018 Series Operation Manual Download Page 20

 

Manual Part Number: 

5-06-7000-80-0 

Manual File: 

012476 Rev. A 

Page 3-20

 

 

 

 

M

AINTENANCE 

&

 

T

ROUBLESHOOTING

 

 

3.1  Replacing the Sample Hose Assembly. 

In  the  event  the  analyzer  displays  an  Error  #3  or  Error  #5  this  may  be  an  indication  the 
sample  hose  needs  replacing.    This  will  occur when  the  integrated  flow  restrictor  becomes 
clogged with oil, debris or sealant.  It can also occur if there is inadequate flow, less than 30 
psig (2 Bar) refrigerant in the vehicle or cylinder. Replacement hoses for both the R-134a and 
R-1234yf  couplers  are  provided  in  the  kit.    Additional  replacements  are  listed  on  the  spare 
parts list in section 

4

 Appendices

 

To replace the Sample Hose Assembly, follow the instructions below: 

 

   

1) 

Disconnect the sample hose from the refrigerant source and Analyzer 

 

   

2) 

Remove the brass restrictor end (with hose attached) from the coupler 

 

   

 

and  discard.  Be  sure  to use  a  backing  wrench as  not  to  damage  the 

 

   

 

coupler. 

 

   

3) 

Check for signs of oil and debris in the coupler. 

 

   

4) 

Use a cleaner which ONLY contains, Tetrachloroethylene and carbon 

 

 

dioxide, follow safety instructions on the can and spray all parts of the 

 

 

coupler with the cleaner to remove any oil. DO NOT soak the part for 

 

 

more than 60 seconds. 

 

   

5) 

Allow coupler to dry. Check coupler for oil once again. Failure to clean 

 

   

 

the oil out of the coupler will result in premature obstruction of the new 

 

   

 

sample hose. 

 

   

6) 

Install the brass end of the new sample hose assembly into the 

 

   

 

coupler and lightly tighten, usually finger tight is sufficient.  

 

3.2  Information Screens 

An 

“Information” icon

  

           or ‘Help’ indication will appear at various points throughout the 

testing  process.  This  button  will  provide  additional  information  or  tips  about  the  command 
screens to help complete your analysis.  

 

 

Summary of Contents for Neutronics LEGEND ID RI-2018 Series

Page 1: ...56 Creamery Way Exton PA 19341 USA Phone 1 610 524 8800 Fax 1 610 524 8807 Email info refrigerantid com www refrigerantid com ACTIVATE SOFTWARE UPDATES BEFORE FIRST TEST www refrigerantid com activate...

Page 2: ...Manual Part Number 5 06 7000 80 0 Manual File 012476 Rev A Page II...

Page 3: ...Manual Part Number 5 06 7000 80 0 Manual File 012476 Rev A Page III LEGEND ID Model RI 2018 Series AUTOMOTIVE REFRIGERANT ANALYZER OPERATION MANUAL...

Page 4: ...USB Cord 1 12 R 1234yf and R 134a Tank Adapter Fittings 1 12 AC Power Adapter 1 13 Control Panel 1 13 Back Panel Connections 1 14 Hard Shell Storage Carrying Case 1 14 2 LEGEND IDTM OPERATION 2 15 2...

Page 5: ...llow the manual can result in serious injury or death Less than 2 grams of refrigerant are vented with each sample This analyzer is designed with sealed heat sources and without sparking components o...

Page 6: ...hout sparking components Ensure adequate ventilation and always take proper precautions when working with refrigerants DO NOT breathe refrigerant and lubricant vapor or mist Exposure may irritate eyes...

Page 7: ...ion regarding the application operation or spare parts please contact the Neutronics Inc Customer Service Department If you have questions or comments we would like to hear from you Neutronics Inc 456...

Page 8: ...by the SAE as a refrigerant mixture that contains 98 0 or greater of R 1234yf or R 134a by weight The instrument is supplied complete with an R 1234yf sample hose an R 134a sample hose R 12 coupler so...

Page 9: ...ant sampled Capable of analyzing R 12 1 4 Flare coupler sold separately Multiple Languages o English German Spanish French Italian Portuguese Chinese Japanese Korean and Russian Easily prints test res...

Page 10: ...therefore there are no serviceable components internal to the instrument and disassembly will void the warranty Legend R 1234yf Sample Hose The 6 5 foot 2 meter R 1234yf Sample Hose is constructed of...

Page 11: ...le hose extensions allow the user to easily connect and disconnect the hose assembly to the analyzer The extension is connected directly to the analyzer and the sample hose connects to the male feral...

Page 12: ...zer R 1234yf Tank Adapter Fitting The R 1234yf Tank Adapter Fitting will provide the user with an adapter to allow connection of the R 1234yf Sample Hose to the LH Acme threads on the R 1234yf cylinde...

Page 13: ...harge the battery when connected to the analyzer NOTE Use of any other power source may cause damage to the unit and void the warranty Control Panel The Control Panel serves as the main user interface...

Page 14: ...structed Keep the sample outlet port free and clear at all times Do not operate near open flame Hard Shell Storage Carrying Case The hard shell storage carrying case is custom fit to the Legend ID It...

Page 15: ...2 2 Power On the Analyzer For use with an SAE J2843 or J3030 certified AC Service Machine connect one end of the provided USB cable to the USB port on the back of the analyzer and connect the other en...

Page 16: ...or cylinder you must select R 134a mode 2 3 Calibration Each time the Legend ID begins a new test cycle it must complete an air calibration The calibration takes 30 seconds and pulls fresh air into t...

Page 17: ...r a few seconds and then press the Test button to begin the test The Testing screen shown in Figure 8 will display 2 5 Viewing the Test Results Upon completion of testing the Legend ID will display Fi...

Page 18: ...rigerant source and press Exit to return to the main screen Figure 11 NOTE In R 134a mode R 12 and R 1234yf are combined into one reading referred to as R 12 R 1234yf 2 6 Understanding Test Results Th...

Page 19: ...the sampled refrigerant is found to be less than 95 presents hydrocarbons or a has large contamination the screen will illuminate Red and CAUTION SHOULD BE TAKEN WHEN HANDLING THIS VEHICLE OR CYLINDE...

Page 20: ...lyzer 2 Remove the brass restrictor end with hose attached from the coupler and discard Be sure to use a backing wrench as not to damage the coupler 3 Check for signs of oil and debris in the coupler...

Page 21: ...ER THE ANALYZER WE WILL NOT BE ABLE TO INFORM YOU OF ANY SOFTWARE UPDATES 3 4 Settings Pressing the Settings button as shown in Figure 18 will provide access to various device settings as shown in Fig...

Page 22: ...hat the white filter is correctly plugged into the rubber grommets Error 4 The unit is beyond the operating temperature range Solution Move the unit to an area where the ambient temperature is within...

Page 23: ...0 Legend R 1234yf Replacement Hose 6 02 6001 56 0 Legend R 134a Hose Kit 6 02 6001 57 0 Legend R 1234yf Hose Kit 4 2 Specifications SAMPLE PARAMETERS Vapor only oil free 500 psig 2 MPa Maximum DETECTE...

Page 24: ...oceedings or other expenses Any description of the goods contained in any documents to which these warranty provisions relate including any quotations or purchase orders relating to the goods being de...

Reviews: