background image

N

2

O Portable Monitor 

Operation 

Instruction 3015-4787 

 

4.

 

Observe that the 

NEW TWA/twa

 screen may appear, depending on 

the configuration of the monitor’s TWA/twa function (refer to 

Section 3.4 for detailed TWA operation). 

5.

 

Observe the monitor starts counting down its warm-up period of 

900 seconds (15 minutes). 

6.

 

At the end of warm up, take the monitor to an area than contains 

fresh air and press the 

ENTER

 button to fill the monitor’s internal 

purge-air bag with clean, fresh air that is of the same temperature 

and humidity as the area being monitored for gas. DO NOT fill the 

purge-air bag in an area that is contaminated with N

2

O gas. 

After the purge-air bag is filled, observe the following as the monitor makes 

gas measurements in the area being sampled:  

 

MEASURE

 indicates when the monitor is actively measuring the 

level of N

2

O gas at the sampling point. The screen to the right shows 

that currently 100 ppm is being measured. The measurement cycle 

runs for 4 minutes.  
A log of up to 200 previous measurements can be viewed using the 

PPM LOG 

function (page 12), or printed as described in 3.15. 

The 

2ndUNIT

 value (Blank, Peak PPM, Continuous TWA, or 8 Hr 

TWA) is user-selected (refer to 

2ndUNIT 

function on page 13). 

 

PURGE

 indicates when the monitor is resetting its infrared detector 

to a baseline of 0 ppm using the air contained in the monitor’s 

internal purge-air bag. The purge cycle runs for 10 seconds. 
The monitor will operate for approximately 40 minutes before the purge-air bag needs refilled. 

When the message "PURGE AIR EMPTY" appears, take the monitor to a 

clean, fresh-air 

environment

 and press the 

ENTER

 button to refill the bag. 

Work with gas alarms and system faults as follows: 

ALARM

 (red) light turns ON if the detected gas level exceeds a programmed alarm point. An audible 

alarm also sounds if the 

AUDALRM

 function is activated (page 11) Pressing the front panel 

SILENCE 

button turns OFF the audible alarm, but the 

ALARM

 light will continue to flash for as 

long as the detected gas level is above the alarm point. The audible alarm will reactivate if the gas-

alarm condition is not cleared within the time period set by the 

SILENCE 

function (page 12). 

FAULT

 (yellow) light turns ON if a fault occurs (see Fault Code list on page 15). An audible alarm 

also sounds if the 

AUDALRM

 function is activated (page 11). Pressing the front panel 

SILENCE 

button turns OFF the audible alarm, but the 

FAULT

 light continues to flash for as long as the fault is 

present. The audible alarm will reactivate if the fault condition is not cleared within the time period 

set by the 

SILENCE 

function (page 12). Both the 

FAULT

 light and audible alarm will automatically 

turn OFF after the cause of the fault has been eliminated. A log of the last 30 fault events can be 

viewed using the monitor’s 

FAULTS 

function (page 12). 

MEASURE          2ndUNIT 

        100ppm   N2O 

PURGE               2ndUNIT 

       100ppm    N2O 

 
   

PURGE AIR EMPTY 

(ENTER) TO FILL 

PURGE AIR EMPTY 

<ENTER> TO FILL 

WARM UP 

900 

Summary of Contents for N2O Portable Monitor

Page 1: ...table Monitor Portable Area Gas Monitor PAGM with Infrared Sensor for N2O Gas Instruction 3015 4787 Operation Maintenance Rev 8 February 2013 Product Leadership Training Service Reliability Patent 6 5...

Page 2: ...N AND ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIG...

Page 3: ...al 6 3 3 Gas Detector Operation 6 3 4 Time Weighted Average TWA Operation 8 3 4 1 Overview 8 3 4 2 Starting a New or Continuing TWA twa 8 3 5 Peak Reading Operation 9 3 6 Leak Detector Operation 9 3 7...

Page 4: ...ta to a Personal Computer 20 3 17 Importing Saved Data Into a Spreadsheet 21 4 MAINTENANCE 22 4 1 Disassembly 22 4 2 Internal Particulate Hydrophobic Filter Replacement 23 5 PARTS AND SERVICE 24 5 1 R...

Page 5: ...ich if not correctly performed or adhered to could result in personal injury or death 1 3 Caution Statements The use of the word CAUTION in this manual denotes a potential hazard associated with the u...

Page 6: ...f the monitor is used in a manner not specified by these instructions Changes or modifications to this monitor will void the warranty 1 5 5 Altitude Limit 6 562 ft 2 000 m 1 5 6 Cleaning Clean the out...

Page 7: ...Overview The Bacharach Nitrous Oxide N2O Portable Monitor is designed to detect measure and display the concentration of N2O gas in the area being monitored To ensure accurate gas measurements the mon...

Page 8: ...to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 562 ft 2 000 m Power DC power pack provides 8 12 hours of operation Power Consumption 15 Watts Front Panel 3 Indicator lights ON Gre...

Page 9: ...ing onto the elbow connector and attach the elbow onto the Gas Sample Port on the front of the monitor as shown in the illustration below The gas sample line can be up to 50 ft 15 2 m in length All tu...

Page 10: ...filter To connect the exhaust line to the exhaust port s PTC fitting firmly push the appropriate tubing into the hole in the center of the connector until it seats in the connector Refer to the figur...

Page 11: ...ir bag The purge cycle runs for 10 seconds The monitor will operate for approximately 40 minutes before the purge air bag needs refilled When the message PURGE AIR EMPTY appears take the monitor to a...

Page 12: ...ll continue to be updated twa lower case letters is continuously calculated over the operating time of the monitor This mode is useful for monitoring an area for a few minutes or more and is the prefe...

Page 13: ...ndby Mode When taken out of standby the monitor is ready for operation without waiting an additional 15 minutes to warm up While in standby the monitor will stop taking gas samples and stops purging i...

Page 14: ...firmware revision level After a brief moment the Warm Up screen is displayed along with the front panel ON light green blinking The monitor requires 15 minutes to warm up from a cold start after whic...

Page 15: ...ugh the displayed data or to change a parameter associated with that function Press ENTER to save any newly entered parameters Press the ESC button to return to the previous screen without saving Note...

Page 16: ...PPM Log can be cleared per Section 3 11 LOG INT Sets the interval at which measurements are logged to memory from 1 to 9999 minutes Factory default is 1 minute The logged measurements can be viewed us...

Page 17: ...ed to appear in the upper right hand corner of the Data Display Screen NONE Blank Peak PPM Peak gas reading Cont twa Time Weighted Average calculated on a continuous basis 8hr TWA Time Weighted Averag...

Page 18: ...rough the fault log In the screens shown to the left record 03 shows that a Sample Flow Fault fault code 0800 occurred on 07 11 05 at 08 17 while record 04 shows that the fault was cleared on 07 11 05...

Page 19: ...minated with gas Take monitor to a clean air area and use the EMTYBAG and FILLBAG functions several times to decontaminate the purge air bag 0200 Gain Set Fault The digipot autotune sequence has faile...

Page 20: ...level the audible alarm will reactivate 3 12 Clearing the PPM Faults Data Up to 200 gas measurements and 30 fault events are stored by the monitor To clear stored data first display the data to be cl...

Page 21: ...isk The purge valve can be opened and closed by pressing the Keypad Left button An asterisk appears on the display when the purge valve is open causing the monitor to draw air from its purge air port...

Page 22: ...d closed Note that an asterisk appears when the purge valve is open causing the monitor to draw air from the purge air bag The Keypad Down button toggles the pump ON and OFF Pressing the ENTER button...

Page 23: ...s using the LOG INT function page 12 2 Interconnect the RS 232 connectors located on the printer and monitor 3 If not already done refer to the printer manual and set the printer s communication param...

Page 24: ...to be made as follows a Click Start click Run In the Run box type hypertrm exe and click OK The New Connection dialog box should appear b Type in a New Connection Name such as N2O Click OK to select...

Page 25: ...eadsheet A text file that was created as described in Section 3 17 can be imported into spreadsheet programs that are capable of importing comma delimited files The following describes how to generate...

Page 26: ...or disassemble the monitor s metal chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack...

Page 27: ...ement filter P N 0007 1654 Medium Phillips head screwdriver Procedure 1 Disassemble the monitor s metal chassis per Section 4 1 2 See illustration below and locate the internal particulate hydrophobic...

Page 28: ...ft 0003 6182 Replacement Pump 3015 4239 Battery for main PCB board 0204 0020 Replacement Bench Kit 3015 5247 Optional Accessories Item Description Part Number IrDA Printer Kits Parts Accessories 120...

Page 29: ...tacting any of the following Bacharach Service Centers United States Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5051 Fax 724 334 5723 Email help mybacharach com Canada...

Page 30: ...Declaration of Conformity N2O Portable Monitor 26 Instruction 3015 4787 6 Declaration of Conformity...

Page 31: ...N2O Portable Monitor Instruction 3015 4787 27 Notes...

Page 32: ...unt Valley Circle New Kensington PA 15068 7074 Toll Free 800 736 4666 Tel 1 724 334 5000 FAX 1 724 334 5001 Website www MyBacharach com E mail help MyBacharach com Printed in U S A Registered Trademar...

Reviews: