1100-2294 Rev 1
25
Depending on the duration the switch is held, a short “TAP” or long “HOLD” will be detected:
To carry out a tap function, tap the relevant switch location for 1 second, until a single
“chirp” is heard, remove wand to confirm a “TAP.”
To carry out a hold function, do not remove the magnetic wand after the first chirp but
continue to hold for >5 seconds
, until a double “chirp” is heard, remove wand to confirm
a “HOLD.”
If either switch is held for >30s, a stuck switch fault will be indicated.
To interact with the instrument without use of the magnetic wand, two internal push button
tactile switches may be used. Remove lid without removing ribbon cable to access. Internal
switches TACT#1 and TACT#2 mirror the functions of MAG#1 and MAG#2.
The function of each switch depends on the current state of the instrument as indicated in the
following table:
State
Switch 1
(Tap)
Switch 1
(Hold)
Switch 2
(Tap)
Switch 2
(Hold)
Warm-up
Enable
Bluetooth
®
Connectivity
-
Disable
Bluetooth
®
Connectivity
-
Normal
Start Zero Calibration
Start Span
Calibration
Low Alarm
Mute Buzzer
Ack. Latched Alarm
High Alarm
Mute Buzzer
Ack. Latched Alarm
Offline
-
-
Fault
Mute Buzzer
Ack. Latched Fault
Negative Gas
Fault
Mute Buzzer
Start Zero
Calibration
Zero Cal. Fault
Acknowledge Fault
-
Span Cal. Fault
-
Acknowledge Fault
4.1.6 Reset System to Factory Default Settings
To reset system to factory defaults, remove lid and hold TACT#1 and TACT#2 simultaneously
for 30 seconds. Instrument will restart to confirm factory reset. Alternatively, see Section 4.2.3.4
“Reset to Factory Defaults”, for instructions on resetting instrument configuration via the MGS-
400 App.