2.8.1 Basic Main Mode
Date and Time:
Date displayed as
mm/dd/yy
, while
time is displayed in a 12-hour format. The date and
time are set per Section 4.8.
Btry
(V):
Battery voltage
Flow
(lpm):
Sample flow rate in either liters/min or
cu.ft./min.
Back
(%):
Measured background gas level in either
percent-by-volume or ppm.
Leak
(%):
Measured natural gas leak concentration at
the current sample flow rate in either percent-by-
volume or ppm.
Leak
(lpm):
Calculated leak rate in either liters/min or
cu.ft./min.*
S
peed
(LO<
––––
|>HI):
Blower speed indicator.
Start / Stop:
Select to Start and Stop a test**
Menu:
Select to display the first of several submenus
(refer to Section 2.9)**
:
Battery status icon (refer to Section 2.10)
[0001]:
Current Test ID number
(A)
or (M):
Automatic or Manual mode (refer to Section
S
--:
Standby
-1-:
Measurement #1 (maximum flow rate)
--2:
Measurement #2 (reduced flow rate)
Message Line:
Various messages appear during
operation describing the current function being
performed, or errors that have occurred (refer to
Section2.19: Message Line on page 38).
Calibration:
Allows user to perform various calibration
functions, including calibrate, verify, view calibration
dates, view amplifier gain values, or erase calibration.
Set Flow Units:
Select to set flow rate display to either
lpm (liters per minute), or cfm (cubic feet per minute).
Select Language
:
Select to set menu language to
English or Russian.
Menu Mode:
Select to set menu to basic mode or
expanded mode.
* “Not Avail.” is displayed when a value cannot be calculated from available information.