IR Emitter Adjustment 

WARNING! 

 

SHOCK HAZARD

When performing the following step, be careful not to come in contact 

with AC voltages when working inside the instrument with the cover off and AC voltage applied.

 

 
15.  Reconnect the AC power cord, and then turn the instrument ON while holding down the front panel 

ENT

 

button to enable the Factory menu 

16.  Wait until the instrument warms up and the Data Display screen appears before proceeding. 

17. Press the 

ENT

 and 

ESC

 buttons simultaneously to access the Main menu. From the Main menu, select 

DIAGNOSTICS > IR EMITTER to display the IR Emitter Parameters screen. 

18.  If the R37 potentiometer is present on your generation of the H25-IR, adjust it clockwise to a reading of

 

450 ±20 mW or 540 ±20 mW 

(R600 model only), corresponding to the version of the part as seen in 

the items required section of this manual containing the part numbers.  If your generation of the H25-IR 
does not have an R37 potentiometer, the IR Emitter power can be adjusted by pressing the 
VOL/RANGE (higher/lower adjustment in small increments) or SETUP/MODE (higher/lower adjustment 
in larger increments) buttons on the front of the unit. 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

 

Figure 3. 

DIGIPOT Adjustment 

19. Press 

ESC

 twice to return to the Main menu. From the Main menu, select FACTORY > DIGIPOT to 

display the DIGIPOT screen. 

IMPORTANT!

 

When performing the following Step, the instrument must be sampling fresh air 

(zero gas) when adjusting the DIGIPOT value.

 

20. Use the 

Up

 and 

Down

 buttons to adjust the DIGIPOT value until the IR VOLT reading is 

4.2000 ±0.0500 volts. Press 

ENT

 to save the new DIGIPOT value.   

  IMPORTANT!

  If adjusting the DIGIPOT does not reach 4.20000±volts then return to step 18 to lower 

the IR Emitter setting.  If the IR Emitter setting is 450±mW adjust the IR Emitter setting 
to 375±mW.  Next, repeat the DIGIPOT adjustment. 
 
 
 
 

 

21.  Turn OFF instrument and remove AC power cord. 

22. Reassemble instrument. 

 

 

 IR EMITTER PARAMETERS

  VOLTS=5.7345  mA=77.5 
    OHMS=75.0 

mW=450.0 

DIGIPOT=220 
IR VOLT=4.2000 

NOTE:

 

Your readings for the other 

parameters on this screen may vary.

 