18
3015-9005 Rev 1
Section 5. Troubleshooting
5.1. Diagnosing Issues
P=Problem
•
C/S=Cause/Solution(s)
P No response to calibration reference
bottle.
C/S Heat Adjustment is set too low or
bottle is empty. Readjust heater (see
“Heater Adjustment” section) or
replace bottle (3015-0864).
No air flow (indicator ball in probe
doesn’t float). Replace filter in probe
tip (3015-0784). Check for proper
pump operation.
Sensor exposed to excessive amounts
of halogen gas. Move probe to clean
atmosphere for several minutes
while sensor purges itself.
Water is in the probe. Turn unit off
and disconnect probe from chassis.
Remove screws and take out chassis.
Look at the underside of the leak
detector and follow the probe to the
pump. Remove this hose from the
pump. Blow clean air (5 psi) into the
probe tip for one or two minutes.
Reassemble unit and replace the
filter (3015-0784).