Instruction 0011-9026
Page 21
accuracy don’t forget to saturate
fi
lter saturator wool packing
as such gas standards are usually supplied “bone dry”.
6.2.2 O
2
FYRITE Fluid Only
Fresh FYRITE
fl
uid will absorb all O
2
from approximately 100 samples containing
10% O
2
. After completing test (as outlined in Section 3.0 Steps 1 through 9) O
2
FY-
RITE Fluid strength can be checked by performing steps outlined below:
1. DO NOT VENT FYRITE, but reabsorb sample gas into FYRITE by inverting until
fl
uid drains into top reservoir, then turn upright to drain
fl
uid into bottom reser-
voir. Repeat this step three (3) more times (four complete inversions total).
2. Hold FYRITE at 45° angle momentarily to allow
fl
uid droplets to drain into bot-
tom reservoir.
3. With FYRITE held upright, permit
fl
uid in column to stabilize a few seconds, then
immediately read % oxygen on scale at the point corresponding to top of
fl
uid col-
umn. If reading increases by more than 1/2 percent O
2
as compared to the initial
reading, replace
fl
uid.
When it is desirable to check
fl
uid strength before taking the FYRITE to a location
where it will be used, it can be conveniently tested on a sample of atmospheric air
which contains 20.9% O
2
.
NOTE:
Make certain wool in
fi
lter saturator tube is moistened
with water when using FYRITE in a non -combustible applica-
tion where the gas sample is not fully saturated with water
vapor.
Formation of a frothy, persistent foam on FYRITE
fl
uid also indi cates need for
fl
uid
replacement. A few “beads” or small air bubbles
fl
oating near the small center bore
wall is not considered foaming. Color of FYRITE O
2
(blue)
fl
uid is not an index of
ability to absorb oxygen although towards the end of
fl
uid life, a greenish cast can be
detected.
NOTE:
If compressed gas standards are used to check FYRITE
accuracy, don’t forget to saturate
fi
lter saturator wool packing
as such gas standards are usually supplied “bone dry”.
6.3 Replacing FYRITE Fluid (Refer to Figs 20, 21, and 22)
CAUTION
Because of the corrosive effect of the FYRITE
fl
uid, always
change FYRITE
fl
uid in immediate vicinity of a sink with run-
ning water available as shown in Figure 18.
Remove 4 screws, metal Top Cap Ring, plastic Top Cap Assembly and Top Gasket.