![Bacharach FYRITE CO2 Operating & Maintenance Manual Download Page 8](http://html1.mh-extra.com/html/bacharach/fyrite-co2/fyrite-co2_operating-and-maintenance-manual_464778008.webp)
Instruction 0011-9026
Page 4
CO
2
FYRITE OPERATION
O
2
FYRITE OPERATION
(RED FLUID)
(BLUE FLUID)
1. Hold upright (Fig. 1) and
away from face. Depress
Plunger Valve (momen-
tarily) to vent FYRITE,
and release.
2. Invert FYRITE (Fig. 2).
Hold at slight angle to
drain
fl
uid into top reser-
voir.
3. Turn upright. Hold FY-
RITE at 45° angle (Fig. 3)
momentarily to allow
fl
uid droplet drainage
into bottom reservoir.
4. Hold FYRITE in upright
(Fig. 4) position and
away from face. Depress
Plunger Valve (momen-
tarily) and release.
Figure 1
Figure 2
Figure 3
Figure 4
1. Hold upright (Fig. 1) and
away from face. Depress
Plunger Valve (momen-
tarily) to vent FYRITE,
and release.
2. Invert FYRITE (Fig. 2) to
absorb O
2
drawn into FY-
RITE whenever Plunger
Valve is depressed). Hold
at slight angle to drain
fl
uid into top reservoir.
3. Turn upright. Hold
FYRITE at 45° angle to
drain
fl
uid into bottom
reservoir. Turn upright.
Repeat sequence of Steps
1, 2, and 3 twice until
fl
uid level does not drop
more that 1/2% when
Plunger Valve is de-
pressed. Turn upright.
Hold FYRITE at 45° an-
gle (Fig. 3) momentarily
until excess
fl
uid droplets
have been drained from
inside surfaces.
4. Hold FYRITE in upright
(Fig. 4) position and away
from face. Note
fl
uid
level in column. Depress
Plunger Valve (momen-
tarily), and release.