background image

CO

2

 Monitor 2800

Instruction 19-9220

3.5  Fault Condition Warning

The unit can warn the user of an internal fault, i.e. a source failure or 

blockage in the Infrared path. When a fault occurs, both LEDs will illu-

minate and the alarm will sound continuously. The following screen will 

display until the unit is switched off.

FAULT

If the fault condition warning is activated at any time, then the unit 

must be returned to Bacharach for service.

3.6  Data Logging

When the CO

2

 Monitor is switched on, it will automatically begin storing 

time, date and CO

2

 levels at 30 second intervals. Optional BACH-COM 

Software allows this interval to be changed (Section 4.1).

Readings are stored with their real time to the nearest second. There is 

storage capacity for more than 1,500 sets of readings. Once the storage 

capacity has been exceeded, the latest readings will overwrite the oldest 

stored data - i.e. first in­, first out.

NOTE:

 Optional BACH-COM Software is available to pro-

duce tables and plots of real time CO

2

 exposure (See Section 

4.1 for part numbers).

3.7  Software Version/Serial Number

While holding the 

f

1

 button down, turn the unit ‘on’ in order to retrieve 

the instrument software version and issue date. Releasing the 

f

1

 button 

will disp­lay the in­strumen­t serial n­umber for five secon­ds.

3.8  Factory Authorized Reset Mode

This function should only be activated when advised by a Bacha-

rach service representative as it may alter calibration.

While holding down 

f

2

 button­, turn­ un­it “on­”. The disp­lay will show... 

 

FACTORY

SETTINGS

Keep 

f

2

 button depressed for 5 more seconds until the display shows...

FACTORY

RESET OK 

6

Summary of Contents for 2800

Page 1: ... CO2 Monitor 2800 Instruction 19 9220 Operation Maintenance Rev 6 June 2004 Product Leadership Training Service Reliability ...

Page 2: ...val of any title defects or at the election of Bacha rach to the replacement of this Product or parts thereof that are defective in title THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHAN TABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGA TION LIA...

Page 3: ...g 2 1 7 Sunlight 2 1 8 Servicing 2 2 0 OPERATION 3 2 1 Reading Mode 3 3 0 MAINTENANCE 4 3 1 Battery Low 4 3 2 Air Calibration 4 3 3 Calibration Using Higher Gas Concentration 5 3 4 Test Mode 5 3 5 Fault Condition Warning 6 3 6 Data Logging 6 3 7 Software Version Serial Number 6 3 8 Factory Authorized Reset Mode 6 4 0 PARTS SERVICE 7 4 1 Parts and Accessories 7 4 2 Bacharach Service Centers 7 ...

Page 4: ...Humidity 10 95 non condensing Response Time 10 seconds to 90 of reading Size 140 L x 65 W x 20 D mm 5 5 x 2 6 x 0 8 in Weight 200 grams 0 44 lbs Case ABS Plastic IP55 NEMA 3R Available Models Standard Instrument 19 3319 Instrument with 110 VAC Charger 19 7101 Instrument with 230 VAC Charger 19 7102 Note that ppm is the standard as supplied With the optional ...

Page 5: ...t and can be hand held or attached to a belt 1 2 Main Features Audible and visual warnings if CO2 exceeds preset levels Easy to operate with user friendly screen display Easy calibration in clean air Functional test mode Fault condition warning The unit is normally supplied complete with a battery charger and oper ating instructions 1 3 Battery The battery should provide 8 to 10 hours of continuou...

Page 6: ...m light D IrDA communication link E Alarm light F Audible alarm location G Test button fl H Battery charging and ear phone socket I Air calibration button f2 J On Off Button 1 6 Cleaning Keep the unit clean by wiping with a soft cloth dampened with a mild detergent solution 1 7 Sunlight The unit should not be left out in direct sunlight or in other areas where excessive heat exists for long period...

Page 7: ...eted as an alarm condition The unit has two preset alarm levels which are normal ly set to indicate the presence of Carbon Dioxide in concentrations above 5 000 ppm or 10 000 ppm factory preset levels The levels can be factory set at customer specified levels or can be reset using the BACH COM Software See Section 4 1 When the CO2 level rises above the low alarm level The display will show the con...

Page 8: ...ide air assumed to be 340 ppm CO2 by pressing and holding down the f2 button for a minimum of 3 seconds It is essential to ensure that the unit is in clean air before attempting an air calibration If this condition is not ensured incorrect readings will be obtained Be careful that your breath does not affect this air calibration procedure A 340 ppm calibration gas cylinder can be used See Section ...

Page 9: ... reading If during the calibration check the instrument reading is outside of the tolerance of the check gas used the instrument can be returned to a Bacharach Service Center see Section 4 2 for recalibration 3 4 Test Mode To activate the TEST MODE switch the unit on and then press and hold down the f1 button Both LEDs will illuminate and the alarm will sound The following screen will be displayed...

Page 10: ...e nearest second There is storage capacity for more than 1 500 sets of readings Once the storage capacity has been exceeded the latest readings will overwrite the oldest stored data i e first in first out NOTE Optional BACH COM Software is available to pro duce tables and plots of real time CO2 exposure See Section 4 1 for part numbers 3 7 Software Version Serial Number While holding the f1 button...

Page 11: ... w Cigarette Lighter Adapter 19 3302 Earphone 19 3303 Protective Boot 19 3304 Stand 19 3307 Carrying Case 19 3311 Replacement Parts Battery Pack 19 3306 110 VAC USA Plug Charger 19 3312 230 VAC European Plug Charger 19 3313 Calibration Accessories Calibration Kit 19 8027 Calibration Cup 19 3305 Calibration Gas 340 ppm CO2 24 1125 2500 ppm CO2 24 1130 7 ...

Page 12: ...ark Bacharach Instruments Int l P O Box 44 39 Lindegade DK 6070 Christiansfeld Denmark Phone 45 74 563171 Fax 45 74 563178 Email mail bacharach dk 8 United States California 7281 Garden Grove Blvd Suite H Garden Grove CA 92841 Phone 714 895 0050 Fax 714 895 7950 Email calservice bacharach inc com New Jersey 7300 Industrial Park Rte 130 Bldg 22 Pennsauken NJ 08110 Phone 856 665 6176 Fax 856 665 666...

Page 13: ...Instruction 19 9220 CO2 Monitor 2800 ...

Page 14: ......

Page 15: ......

Page 16: ...Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 800 736 4666 Website www bacharach inc com E mail help bacharach inc com Printed in U S A ...

Reviews: