Instruction 24-9435
15
Fyrite Tech
4.2 Warm-Up Screens
.
As soon as the analyzer is turned ON, a series of Warm-Up Screens are
dis played. These screens show the analyzer’s model number (Tech E050 or
Tech E060), the unit’s software revision, and fi nally the remaining warm-
up time counted down from 60 seconds. At the end of warm-up, the ana-
lyzer dis plays the Fuel Screen (refer to Section 4.3).
Note:
If any errors were de tect ed during warm-up, the Sen-
sor Error Screen (refer to Section 4.11) is displayed.
4.3 Fuel Screen
The Fuel Screen is displayed after warm-up, and is where an operator
chooses which fuel is to be used by the analyzer to calculate effi ciency and
stack loss. Press ing the
▲
or
▼
button scrolls the dis play through the fuel
codes. After se lect ing the fuel, press
ENTER
to select the dis played fuel
and place the analyzer into its Run Mode. Note that the fuel code will be
dis played in the top right corner of all screens while in the Run Mode.
Fuel Codes:
F1 = Natural Gas
F2 = LPG
F3 = Light Oil
F4 = City Gas
Front Panel Button Functions:
▲
– Displays next fuel code
▼
– Displays previous fuel code
ENTER
– Go to Run Mode (refer to Section 4.13)
HOLD
– No effect
LIGHT
– Toggles backlight ON/OFF
I/O
– Turns analyzer OFF (with a 5 second delay)