Select 'Set Zero Level' by rotating the black button and press to confirm.
Rotate the black know to change 'No' to 'Yes' and press to confirm your choise. When choosing 'No', you return to
the previous menu. When pressing 'Yes' you access the screen below:
The sample counter moves from 1 to 16 and then the next screen is shown:
Rotate the black know to dial 7 and press to confirm the zero setting. The display shows 'Zero Level successfully
changed' to confirm the setting. If the black button is pressed without dialing7, the zero level is not stored.
In case the sensor is not in the correct position (sensor too high or too low) to allow a correct full range
measurement, the display will indicate an error first and the position of the sensor must be modified after
which the procedure must be repeated. Examples of possible alarm messages in case of false sensor
position are shown below.