Advanced operation:
1. Time mode
• System enters time mode when:
1. Power is on;
2. No key is pressed for more than one minute.
• For normal status:
1. Display time in a sequence of year, month, date, week, hour, minute with a flickering " : " stands
for second.
2. Press "Talk" to report current time.
3. Press hold "Talk" for 2 seconds to activate or disable the talking function - During measurement,
you still hear "didi" no matter whether it is in talking or non-talking mode
4. Press "Memo" to retrieve the recorded temperature and Press Hold "Memo" for 2 seconds to
initiate setup mode.
5. Press "C / °F" to select Celsius or Fahrenheit
• For setup mode of the hour and date:
1. Press¡¾ Talk ¡¿to shift between setting of hour, minute, year, month and date.
2. Press and hold "Memo" for approximately 3 seconds until you hear the signal, the hour will start
flashing, next pressing "Memo" you change one by one the hour. Later press "Talk" and the
minutes start flashing. Change the minutes one by one pressing the "Memo" button. If you press
the "Talk" button again, you'll enter the year setting mode and you change it by pressing
"Memo". In such a way you'll set the next month and day.
3. Press "Memo" subsequently to resume to normal status.
* Unit returns to time mode automatically 1 minute after last key pressed.
2. Memory search mode
• Search mode can be activated by pressing "Memo" in normal time mode.
• "-" is displayed if no data are stored;
• Search number is shown for stored data available. Maximum storage is 30 sets of data;
• Press "Memo" to go to the next previously made measurement.
• Unit returns to time mode automatically for 5 seconds after last key pressed.
Recommendations:
1. Measure only the front of the head exposed to the environment. Anything covering the area to be
measured (hair, hat, wig, bandages) would insulate the area, resulting in falsely high readings.
2. Wait about 30 seconds before measuring the same patient again to avoid excessive cooling of the
skin.
3. When exposed to changing room temperature, the device should at least remain in the same room in
which the temperature will be measured for 15 minutes prior to the measurement
4. Always cover the thermometer with the protective cap supplied.
Maintenance, storage and calibration:
Maintenance:
1. Remove any stains on the body with a soft and dry cloth.
2. Don't wash with water or detergent containing abrasive or benzene.
Storage:
1. Always put the protective cap over the probe if the unit is not in use.
2. Don't leave the thermometer exposed to or at high risk of direct sunshine, high temperature, damp,
fire, flame, vibration or impact.
3. Remove the battery in the case of long time without use.
CALIBRATION :
This thermometer is calibrated at the time of manufacture. If the thermometer is used according to the
instructions, periodic re-calibration is not required. If at any time you question the accuracy of the
measurement, please contact the retailer immediately
* Don't attempt to modify or reassemble the thermometer.