background image

 

Page 2 

Page 3 

  BBN01-IM-111109 

NOTE: For directions on cleaning and disinfecting the parts of the nebulizer, refer to the 
“CLEANING” and “DISINFECTION” sections in this instruction manual. Always check to 
ensure the nebulizer is fitted with a filter before use and also ensure the filter is not 
clogged (dirty) before operating the nebulizer. 

 

This nebulizer is a medical device. Only use medications as prescribed and instructed by your doctor. 

When using electrical appliances, basic safety precautions should always be followed including the following: 

 DANGER –To reduce the risk of electric shock 

 

Do not use the nebulizer while bathing. 
Do not place or store the nebulizer where it can fall or be pulled into a tub or sink. 
Do not place in or drop into water or other liquids. 
Do not reach for the nebulizer that has fallen into water. If so ,unplug immediately. 

 Warning –To reduce the risk of injury 

 

  NEVER leave the nebulizer unattended when plugged in. 

  DO NOT leave the nebulizer unattended with people who require supervision. 

  Use the nebulizer ONLY for its intended use as described in this manual. Use ONLY with the supplied nebulizer accessories or accessories 

recommended by Babybelle. 

  Close supervision is necessary when any appliance is used near children or incapacitated persons. 

  NEVER operate the nebulizer if it has a damaged cord or plug, return the nebulizer to the nearest Authorized Service Centre for examination, 

repair or mechanical adjustment. 

  NEVER touch the nebulizer when your hands are wet or moist, as the nebulizer is not protected against water penetration. DO NOT use the 

nebulizer while you are in the bath or having a shower. Never immerse the nebulizer in water. In case this accidentally occurs, always first 
disconnect the power supply plug from the wall socket before reaching for the unit. After removing the nebulizer from water, do not use it again. 
Return the nebulizer to the Authorized Service Centre and have the nebulizer checked.   

  The nebulizer does not contain any user serviceable parts and does not require oil. Do not attempt to lubricate any internal parts. Any servicing 

should be performed by the authorized service representative. 

  Always place the nebulizer on a plain and stable surface during operation. 

  DO NOT expose the nebulizer to sources of heat such as radiator or sun shining through a window. 

  DO NOT let the cord contact any hot surface.   

  DO NOT use an extension cord. 

  NEVER block the air vents on the left/right side of the nebulizer or place it where the vents may be blocked. Make sure there is at least 15cm (6 

in) clearance on all sides of the unit during the operation. Also keep the air vents free of lint, hair, etc.   

  NEVER use the nebulizer while sleeping or drowsy. 

  NEVER drop or insert any object into the open holes of the nebulizer. 

  DO NOT use outdoors. The nebulizer is for indoor use only. 

  DO NOT use the nebulizer in an oxygen enriched environment.   

  DO NOT operate the nebulizer while it is enclosed in a carrying case or bag. 

  The nebulizer has been designed for intermittent use 30 minutes On, 10 minutes Off. 

  Unplug in the nebulizer from outlet when not in use, before cleaning or changing accessories. 

  For medication, please follow the instructions of your doctor or respiratory therapist.   

  ALWAYS turn the nebulizer off before filling the nebulizer cup with medication. 

  Although the device fulfills provisions of the EMC (Electromagnetic Compatibility) directive, using it in close vicinity of other devices should be 

avoided. 

  If the device abruptly stops operating, check that the power cord connection is secure. Also see note in “SERVICE AND CUSTOMER CARE” 

section. 

  Specification regarding output, output rate and particle size may vary depending on the type of medication used (especially for suspensions and 

high viscosity medications). See drug supplies data sheets for further details.     

 

Warning: 1.  The child must not be left alone with the device. 

2.   This product is not suitable for use in anaesthetic breathing systems or lung ventilator breathing systems. 

3.   Performance information provided in manuals of this product may not apply to drugs supplied in suspension or high 

viscosity form.    In such case, information should be sought from the drug supplier. 

4.   The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and 

cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement parts for internal components, may result 

in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYSTEM. 

 

 

 

 

 

 

 

 

 

 

This product is only suitable to alleviate the pain suffered from upper respiratory disease. 

 

Nebulizer Accessory Set 

  Nebulization treatment : Following prescription from doctor, use bronchus soothing dose or phlegm reducing dose. 

  Nasal cavity moistening : To facilitate the flowing out the nasal mucus.   

 

Suction Bottle Set 

  Design for safe and moderate use, solve the snot problem of baby. 

  Suck out the snot from the nasal cavity of baby before feeding.   

  To avoid swolling the nasal mucus and causing pharyngitis and lower respiratory tract infection, suck out the nasal mucus before sleeping.   

 

Nasal Spray Bottle Set 

  Use saline solution to spray the nasal cavity to clean out the dust and restore the normal functioning of nasal mucosa. 

  When changing diapers, use clean warm water to spray wash the skin to avoid rashes caused by wiping using tissues.     

    Use saline solution to clean the wounds and help to heal the wounds.

 

 

 

 

 

  DO NOT plug the nebulizer in before making sure that the electrical rating shown on the rating plate on the bottom of the nebulizer is the same as 

the mains. 

 

The lead of this product is fitted with a non-rewireable plug. In the event that the plug does not fit for the power sockets, an adaptor must be used

 

 

  Unpack all parts of the nebulizer from the packing and remove all plastic bags. 

  Prior to use, clean and disinfect the nebulizer kit and applicable inhalation 

attachments before using it. It is also recommended that the user repeat the process if the nebulizer has not been used for more than one week. 

 
 
 
 

 
 

 

   
     
   

 

1.  Check the nebulizer cup if there is any damage before each use (the nebulizer cup is disposable and should be replaced regularly). 
2.  Place the nebulizer on a firm flat surface near an electrical outlet and in an area where you can sit comfortably and the controls can be easily 

reached during the treatment. 

3.  Make sure the power switch is turned off before you connect it to power outlet. 
4.  Wash your hands before preparing the medication. 
5.  Connect one end of the Air Tubing to the Air Outlet Connector ”A2” (Below the mark A on the nebulizer and ensure tubing is securely attached to 

the nebulizer to obtain proper operation/ treatment.) 

 

 

6.  Rotate the nebulizer cap counter-clockwise from the nebulizer cup and lift the nebulizer cap out of cup . 
7.  With the conical in the nebulizer cup, add the correct amount of prescribed medication to the nebulizer cup and reassemble the nebulizer . 
8.  Connect the remaining end of the air tubing to the bottom of the nebulizer cup. 
9.  Attach one of the following inhalation attachments to the nebulizer cup. 

 The 

mouthpiece 

”B1”. 

 Adult 

mask 

”H” 

  Child mask ”G” 

12.   Switch on the nebulizer by pressing the On/Off switch “A1”. 
 

 
 
 
 
 

INHALATION (NEBULIZER KIT

)

BEFORE PLUGGING IN THE NEBULIZER

NOTE: Do not operate the nebulizer without a filter or when the filter is dirty. Operating 
the nebulizer with a dirty filter can impair proper operation and cause damage that is not 
covered by the warranty. Refer to the” REPLACING THE FILTER” section for details. 

NOTE: Make sure that the Air Tubing is connect to Air Outlet Connector marked A 

INTENDED FOR USAGE 

PREPARING THE NEBULIZER FOR USE 

SAFETY INFORMATION

 

Reviews: