MI 1005 User Manual V1.19
08. August 2013
2. Intended
Use
The
MI 1005
is an appliance especially developed for use in medical technology. The purpose of the
MI 1005
is to connect, in compliance with the relevant norms for electrical safety (IEC 60601-1) and
EMC (IEC 60601-1-2), a medical appliance or system equipped with a network interface to a
non-medical appliance or system (Single PC or local network).
Examples for the use of the
MI 1005
:
x
Connection to the practice or clinic networks with the PC of a prolonged EEG system
x
Connection to an analysis computer in the doctor’s or consulting room with an EMG
measuring station
Point d. of Chapter 1 - General Safety Advice - is to be observed for this.
3. Installation
Only qualified and trained personnel should connect the
MI 1005
. Please instruct your network
systems administrator or your medicine technology department for this purpose. The installation of the
MI 1005
is done by means of the two pluggable connections (RJ-45-sockets). To do this, plug the
network connection cable into the socket of the
MI 1005
, until the lock engages with an audible ‚click’.
The installation of the
MI 1005
should be done near the medical appliance/system. The safety advice
in Chapter 1.2 – Connection - is to be observed!
4. Maintenance
For maintaining a safe technical condition, an annual inspection is recommended, which includes the
following checks:
1. Examination for external damage (housing, network connections, legibility of labeling,
dirt, etc.) of the
MI 1005
.
2. Examination for substances which may have penetrated, in particular moisture.
3. Function
test.
4. Isolation test. The
isolation resistance of the MI 1005 should not exceed 20 MOhm.
For testing, at both connections all data cables as well as the screen should always
be shut down briefly. The isolation resistance should be measured between signal
entry and signal exit.
5. Availability and sufficiency of the documents.
Should any deficiency be established during a test, the MI 1005 should not be used any longer,
as the safety of the patients and the operator cannot be guaranteed.
In case of doubt please contact your supplier or the manufacturer.
5. Cleaning
Use a dry or a moderately moisturized cloth to clean the
MI 1005
. It must be noted that no fluids or
moisture will penetrate into the device.