Page 4 of 10 

MI 1005 User Manual V1.19 

08. August 2013 

2. Intended 

Use 

The 

MI 1005

 is an appliance especially developed for use in medical technology. The purpose of the 

MI 1005

 is to connect, in compliance with the relevant norms for electrical safety (IEC 60601-1) and 

EMC (IEC 60601-1-2), a medical appliance or system equipped with a network interface to a  
non-medical appliance or system (Single PC or local network). 
 
Examples for the use of the 

MI 1005

 : 

 

x

 

Connection to the practice or clinic networks with the PC of a prolonged EEG system 

x

 

Connection to an analysis computer in the doctor’s or consulting room with an EMG 

measuring station 

 
Point d. of Chapter 1 - General Safety Advice - is to be observed for this. 
 

3. Installation 

Only qualified and trained personnel should connect the 

MI 1005

. Please instruct your network 

systems administrator or your medicine technology department for this purpose. The installation of the 

MI 1005

 is done by means of the two pluggable connections (RJ-45-sockets). To do this, plug the 

network connection cable into the socket of the 

MI 1005

, until the lock engages with an audible ‚click’. 

The installation of the 

MI 1005

 should be done near the medical appliance/system. The safety advice 

in Chapter 1.2 – Connection - is to be observed!  
 

4. Maintenance 

For maintaining a safe technical condition, an annual inspection is recommended, which includes the 
following checks: 
 

1.  Examination for external damage (housing, network connections, legibility of labeling, 

dirt, etc.) of the 

MI 1005

 . 

2.  Examination for substances which may have penetrated, in particular moisture.  
3. Function 

test. 

4.  Isolation test. The 

isolation resistance of the MI 1005 should not exceed 20 MOhm. 

For testing, at both connections all data cables as well as the screen should always 
be shut down briefly. The isolation resistance should be measured between signal 
entry and signal exit.

 

5.  Availability and sufficiency of the documents. 

 
Should any deficiency be established during a test, the MI 1005 should not be used any longer, 
as the safety of the patients and the operator cannot be guaranteed.  

 
In case of doubt please contact your supplier or the manufacturer. 
 

5. Cleaning 

Use a dry or a moderately moisturized cloth to clean the 

MI 1005

. It must be noted that no fluids or 

moisture will penetrate into the device. 
 

Summary of Contents for MI 1005

Page 1: ...Page 1 of 10 User Manual Medical Isolator MI 1005 Galvanic Ethernet Isolator for medical electrical appliances in accordance with IEC EN 60601 1 and IEC EN 60601 1 2...

Page 2: ...2 Connection 3 1 3 Environmental Protection 3 1 4 Meaning of used symbols 3 2 Intended Use 4 3 Installation 4 4 Maintenance 4 5 Cleaning 4 6 Declaration of Conformity 5 7 Technical Specification 7 7...

Page 3: ...e MI 1005 should also not be operated in close proximity to devices that emit electromagnetically radiation to avoid electromagnetically interferences like for example mobile phones Potential interfer...

Page 4: ...e socket of the MI 1005 until the lock engages with an audible click The installation of the MI 1005 should be done near the medical appliance system The safety advice in Chapter 1 2 Connection is to...

Page 5: ...Page 5 of 10 6 Declaration of Conformity...

Page 6: ...Page 6 of 10 MI 1005 User Manual V1 19 08 August 2013...

Page 7: ...ction of the user parts against electric shock Not applicable Degree of protection against discharge from defibrillators Not applicable Protection during use in the presence of combustible compounds o...

Page 8: ...but in no event later than two 2 weeks following delivery and return the Products to Company at Company s expense In the event of any apparent shipping and freight damages of the Products Customer sh...

Page 9: ...LIABLE FOR ANY INDIRECT INCIDENTAL PUNITIVE SPECIAL OR CONSEQUENTIAL DAMAGES INCLUDING BUT NOT LIMITED TO DAMAGES FOR LOSS OF PROFITS REVENUE GOODWILL OR USE INCURRED BY CUSTOMER OR ANY THIRD PARTY W...

Page 10: ...t 2013 10 Manufacturer Contact For queries and problems please contact Manufacturer Baaske Medical GmbH Co KG Bacmeisterstr 3 32312 L bbecke Germany Phone 49 5741 236027 0 Fax 49 5741 236027 99 E Mail...

Reviews: