7 Test Procedure
32
DHV0017 b2 EN
©
b2 electronic
GmbH
Step
Procedure
(Run Single Test)
Example
Set “Gap”
& “Rise”
To set the different options, use “Up / Down” arrows to
place the cursor:
•
Electrode Gap: between 0.5
–
10 mm
•
Count: max. 10 counts
•
Rise Rate: 0.5
–
10 kV/s
•
max. Voltage: depends on the max. Voltage of
the BA unit
To toggle the value use Left / Right arrows.
To start test, set the cursor on option “Start Test”
Sample Number
Use the Up / Down arrows to enter Sample Number.
Up/down: Edit Character
Left: Delete Character
Right: Next Character
Click “OK” to continue.
Test Active
The test cycle begins.
The BA automatically detects breakdowns records
them as they occur in the sequence.
During the test the cursor is
automatically on the “Stop”
field. To interrupt the test in progress,
Click “OK” and the test will immediately terminate.
Test Result /
End Test
The test cycle ends.
The summary of the test results appear on the screen
and the test report is printed.
The test has ended.
To return to the main menu, click
„
OK
“.