WF-2000
Instructions and warranty
Solution: It is necessary to consider that the thermometer has to be
in the room where the measurement is performed about half an hour
before performing the measurement, while the person has to be in
the room at least 10 minutes before. (Room temperature has to range
b10°C and +40°C / between 50°F and 104°F)
Drawing 4
Problem: Room temperature is not included between +10°C and
+40°C / between 50°F and 104°F.
Solution: It is necessary to consider that the thermometer has to be
in the room where the measurement is performed about half an hour
before performing the measurement, while the person has to be in
the room at least 10 minutes before. (Room temperature has to range
b10°C and +40°C / between 50°F and 104°F).
Drawing 5
Problem: Error 5~9, the device does not work correctly.
Solution: Remove the battery, await 1 minute and then place it again
in the proper compartment.
Should the error persist, contact your dealer.
Drawing 6
Problem: Temperature exceeds 42.2 2°C – 108°F.
Solution: Verify the integrity of the probe and perform a new meas-
urement.
Drawing 7
Problem: Temperature is lower than 22°C – 71.6°F.
Solution: Verify that the probe is clean and perform a new measure-
ment.
Drawing 8
Problem: The device blocks on self-test.
Solution: Replace the battery
Summary of Contents for WF-2000
Page 1: ...www sotmarket ru B Well WF 2000 8 800 775 98 98...
Page 2: ...INFRARED FOREHEAD THERMOMETER WF 2000 ENGLISH ITALIANO FRAN AIS ESPA OL PORTUGU S DEUTSCH...
Page 15: ...1 C F 25 758 EN 1 3PN...
Page 16: ...WF 2000 1 2 3 4 5 6 7 8 9...
Page 17: ...10 11 37 37 5 0 5 37 1 36 5...
Page 18: ...WF 2000 10 35 2 36 7 35 7 37 3 35 7 37 3 36 2 37 7...
Page 19: ...0 2 1 discharged battery 1 2 3 4 5 1 2 3 ON MEM...
Page 20: ...WF 2000 HEAD C C 4 SCAN 5 5 30 37 5 C 6 SCAN 7 1 1 ON MEM HEAD C...
Page 21: ...2 ON MEM SCAN HEAD 3 4 1 HEAD 1 SCAN ON MEM F 2 SCAN ON MEM 25 HEAD 22 C 42 2 C 71 6 F 108 0 F...
Page 22: ...WF 2000 25 ON MEM M 1 25 1 1 ON MEM 30 1 HEAD HEAD...
Page 23: ...2 3 10 10 40 50 F 104 F 4 10 40 50 F 104 F 10 10 40 50 F 104 F 5 5 9 1 6 42 22 108 F...
Page 24: ...WF 2000 7 22 71 6 F 8 1 2006 92 30 01 2006 BF 1 CR2032LI 1...
Page 25: ...22 80 7 6 F 176 6 F 0 1 10 40 10 50 22 2 0 22 42 2 0 3 42 2 2 0 12 c...
Page 26: ...WF 2000 Well 19 01 98 55 1 2 09 01 1996 5 2 Well 10 Well...
Page 30: ...200 200 200 200 200 200 200 200 WF 2000 200 200 200 200 WF 2000...
Page 81: ......
Page 82: ......
Page 83: ...B Well Limited 758 Great Cambridge Road En eld Middlesex EN1 3PN United Kingdom www b well ru...
Page 84: ...2 B Well WF 2000...