Installation 

 
 

Above ground Installation 
 

It is recommended to lag bolt the ramp in place to keep the scale from moving during 
use. First position the scale and ramp in the location that you intend on using the scale, 
ensure that the scale is away from general traffic and that the chosen area is protected 
from water, misc. debris, wind, and or vibration.  
 
Once the location of the scale has been chosen, remove the scale from the ramp taking 
care not to move the ramp. Using a black Sharpie mark the small bolt holes located on 
the foot retaining plate shown in the image below. 
 

 

 
Once these holes have been marked, remove the ramp and drill lag bolt holes where 
the black Sharpie marks are located on the floor. Once the holes have been drilled 
position the ramp back in place and secure it to the floor using lag bolts. 
 
 
 
 
 

Foot retaining plate 
mounting bolt holes. 

Foot retaining plate 
load cell feet holes. 

Summary of Contents for Above-ground

Page 1: ...Instruction Manual Wheelchair Patient Scales 7 23 Rev V2023 0...

Page 2: ...irst 3 2 Introduction Specifications 4 3 General Component Identification 5 4 Installation Above Ground 6 7 5 Installation In ground 8 9 6 Indicator Termination Calibration 10 7 Replacement Parts 11 8...

Page 3: ...Do not use this equipment for anything other than its intended use Ensure equipment is plugged into stable power sources Keep equipment away from high voltage vibration and high draft areas Keep equip...

Page 4: ...o The B TEK Patient scales are perfect for applications where weighing patients on or off wheelchairs is done in high volume o The low profile above ground with ramp or inground scales makes weighing...

Page 5: ...General Component Identification Above ground A Base platform B Handrail C Ramp General Component Identification Inground A Base Platform B Handrail C Pit Frame A C B C B A...

Page 6: ...on Once the location of the scale has been chosen remove the scale from the ramp taking care not to move the ramp Using a black Sharpie mark the small bolt holes located on the foot retaining plate sh...

Page 7: ...low Level the scale by adjusting the load cell feet through the access holes using a flathead screwdriver shown in the image below The scale is now ready to be connected to the indicator and calibrate...

Page 8: ...ensure the scale is in a place that will not interfere with existing building wiring plumbing etc and at the same time ensure that the scale is away from general traffic and that the chosen area is pr...

Page 9: ...rest on the cable after its installed After the scale has been installed take a flathead screwdriver and adjust the scale base feet so that it is level and even with the floor Note If the scale is ins...

Page 10: ...r patients but also gives you peace of mind knowing your scale is in good working order and working as it is intended to work Termination procedures vary from indicator to indicator If you are using a...

Page 11: ...100092 Load Cell Both 864 000008 J Box with Summing Board Both 911 000032 Load Cell Foot Both 200 84690 Handrail Assy Both 913 1224 Replacement Ramp Above Ground 48 x48 913 1223 Replacement Ramp Abov...

Page 12: ...B TEK Scales LLC 1510 Metric Ave S W Canton OH 44706 Tel 330 471 8900 Fax 330 471 8909 B TEK com...