C100 Series Technical Manual 

 

 

- 18 - 

 

 

8. CALIBRATION 

 

Note: Capacity and division size for local base can NOT be changed 

 

  Turn on the scale and press 

 during the self test. 

 

  Use the numeric key to enter the password 

 
Default password 

0000

 

           Press 

 to confirm 

 

  Display will be show 

 
Select Local or Remote scale by pressing 

  

      Press 

 to confirm. 

 

  Display will be show 

 

If necessary, press 

 to select the weighing unit 

kg

 or 

lb. 

     Press 

 to confirm. 

 

  Display will be show 

 
Ensure the platform is empty and wait for stable indicator. 
Press 

 to confirm. 

 

  Display will be show 

 
Set weight value will be required 
Enter the value by using numeric keys 

           Press 

 to confirm. 

 

  Display will be show     

 
Place the calibration weight on the platform and  
wait few seconds for display to be stable. 
Press 

 to confirm. 

 

  After the calibration the scale will start a self test, remove the weight 

during that time and display will return to weighing mode. 

 

  If display shows any error message or incorrect measurement, repeat the 

calibration process again. 

Pi n  

Pi n  

- - - - 

tech 

Local 

remote  

tech 

unit 

unload 

Load 

sel 

000000 

000005 

Summary of Contents for 899-300000

Page 1: ...C100 Counting Scale Technical Manual V2 40 ...

Page 2: ......

Page 3: ...le to determine the Unit Weight 11 Enter a known Weight 12 Automatic update of Unit Weight 12 Check weighing of count pre set 12 Setting the checking limits 12 PLU Product Look Up 13 Storing PLU 13 Entering Description manually 14 Recalling PLU 14 6 PARAMETERS 15 Keys into the parameter 15 Parameter settings 16 7 RS 232 OUT PUT 17 8 CALIBRATION 18 Linear 19 9 TECHNICAL PARAMETER 20 10 ERROR CODES ...

Page 4: ...lb 00005 kg 0001 kg 0002 kg 0005 kg Units of Measure lb kg Interface Bi directional RS 232 Interface Stabilization Time 2 Seconds Operating Temperature 32 F 104 F Power supply AC Adaptor 9 V 800 mA Battery 6V4AH Calibration Automatic external Display 3 x 6 digit LCD digital display with white LED back light Housing Indicator ABS Plastic Stainless Steel pan Platter size 8 9 x 11 8 Overall dimension...

Page 5: ... air conditioning vents Avoid unsuitable tables The table or floor must be rigid and not vibrate Do not place near vibrating machinery Avoid unstable power sources Do not use near large users of electricity such as welding equipment or large motors Avoid high humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air movement such a...

Page 6: ... is not damaged and that all accessories are included If equipped remove the shipping safety screw Indicated by the red label on the bottom of the C100 Accessories 1 C100 2 Power Adaptor 3 Stainless steel pan 4 18 pre made remote base cable 5 Product manual 6 Owners Manual Keep the packaging material for future use ...

Page 7: ...uld damage the load cell inside Check the water mark If bubble is not centered adjust the leveling feet until centered Check the level when you change the location Attach the power supply cable to the connector on the bottom right side of the scale base Plug in the power supply Turn on the Scale The power switch is located at the right side of the scale base The scale will show the model number in...

Page 8: ... Press key to change between remote base and local base Remote Scale Connection The cable from the remote base goes to a 9 pin male DB9 connecter on the left side of the C100 with following connections Note The sense wires connections of a six wire load cell are not used but can be connected to the respective Excitation pins The remote scale provides 5v excitation Pin Connection Color on Premade c...

Page 9: ... the total weight and displays the result as a net weight Manually enter the weight of sample also changes the unit from lb to kg if they are enabled To enter the numbers of items used as the sample unit weight Add the current count data aggregated Also recall the memory if pressed when balance empty Can add up to 99 values or until it reaches the maximum displayable digits Enter to store and reca...

Page 10: ...ghing Mode Unit Weight Display Smpl of sample is very low U Wt Unit weight is below the minimum weight M Data entered into the memory Local Remote Active Scale in use Count Display CkPcs Active in Counting Mode Wt Active in Weighing Mode High Check Result above the high limit OK Check Result with in the limit Low Check Result below the low limit Charge Status of the battery charging ...

Page 11: ...ighing functions are same for both the scales local or remote The number of weighing divisions may be less on the remote scale dependant on the total capacity of the load cell s used 3 Zero To zero the scale at any time press the key 4 Tare The weight of any container can be tared by pressing the button so that with subsequent weighings the net weight of the object being weighed is always displaye...

Page 12: ...f a tare value of 103g is entered into the 60Kg scale with 5g readability then the display will be shown 105g 5 Accumulation The C100 can totalize weight values or count quantities Manual Accumulation The values weight and count shown on display can be add to the memory by pressing key Set the parameter F1 off print au off Place the goods to be weighed Wait few seconds for display stability then p...

Page 13: ... for display stability and a control beep Unload the item from the platform the weighing value is added into the memory The scale must return to come zero or negative number before adding another samples 99 entries can be stored or until the capacity of the weight display is exceeded 6 Parts Counting In order to do parts counting it is necessary to know the average weight of the items to be counte...

Page 14: ...count will increase If the scale is not stable the calculation will not be completed If the weight is below zero Count display will show negative count Enter a known Unit Weight If an a known unit weight it can enter by using the numeric keys Enter the value of unit weight by numeric keys Press during the unit display flashing If in the weight display as kg unit is active the average piece weight ...

Page 15: ...lect counting Press key the active high count limit will be shown Use numeric key to enter desired and if necessary press to clear Press key the active low count limit will be shown Use numeric key to enter desired and if necessary press to clear Press key to return weighing mode For check limits just one limit value can set If both values are deleted the check mode is deactivated The beep sound w...

Page 16: ...ght to be stored can be either taken from a weighing in process or by enter manually Press display will be shown Press numeric key exp 27 Press Currently stored text will appear The first digit is flashing it can Change by using numeric keys If necessary delete additional text By pressing Continue to enter text until description is complete max 12 characters Use key number selection to left Use ke...

Page 17: ...S 8 T U V 9 W X Y Z 0 _ _ space Note that this method is only used where alpha numeric data is permitted This is used for the Description field and the User ID number Scale ID number in the parameters section Recalling PLU To recall the PLU values the user should first select either local or remote scale the tare value stored will be specific to the scale selected Press display will be shown Press...

Page 18: ...isplay will be show Select the Menu Block Press key it can choose menu block one by one Enter the Selected Menu Press key to confirm what the displays shows Select the Sub Menu Press key it can choose the sub menu block one by one Return to Weighing Mode Press key to escape from the menu and exit to weighing mode F1 off ...

Page 19: ...isable 3 Scale will turn off after 3 minutes 5 Scale will turn off after 5 minutes 15 Scale will turn off after 15 minutes 3 0 Scale will turn off after 30 minutes F2 prt P mode Print Au on Data out put accumulation after unloading the balance Au off Data out put accumulation after by pressing P cont RS 232 data output continuously Ser re RS 232 data output weight only P baud B 600 Set the require...

Page 20: ... 9600 Baud 8 data bits No Parity Connector 9 pin socket From C100 Pin 2 Output Pin 3 Input Pin 5 Signal Ground To LP50 Pin 2 Input Pin 3 Output Pin 5 Signal Ground Connection to the LP 50 printer Set Parameter F2 to 9600 8n1 LP50 Set LP50 Dip SW 1 2 and 4 ON Sample Print LOCAL SCALE ID 123ABC NAME Text 12 456 kg NET 1 1234 g U W 11 PCS TOTAL 49 824 kg TW 44 TPC 4 No ...

Page 21: ... will be show Ensure the platform is empty and wait for stable indicator Press to confirm Display will be show Set weight value will be required Enter the value by using numeric keys Press to confirm Display will be show Place the calibration weight on the platform and wait few seconds for display to be stable Press to confirm After the calibration the scale will start a self test remove the weigh...

Page 22: ...ator Press to confirm Display will be show Apply the weight on the display indicated in chart below wait for stable indication Then press to confirm Display will be show Apply the weight on the display indicated in chart below wait for stable indication Then press to confirm Model Unload Load x Load xx C100 6lb 0 lb 2 lb 6 lb C100 12lb 0 lb 4 lb 12 lb C100 30lb 0 lb 10 lb 30 lb C100 60lb 0 lb 20 l...

Page 23: ...wn Pi n Enter the password Default password is 0000 and press to confirm Pi n Select the scale by pressing which should be configured and press to confirm Tech local remote Use the key to select the weighing unit kg lb and press to confirm Tech unit cnt Use the key to scroll to select individual menu Confirm selected menu by pressing Press key escape from the menu and exit to weighing mode ...

Page 24: ...scale division 2 5 10 20 50 Di v I nc 5 Set division I nc 10 I nc 20 I nc 50 A 2t Azn 0 5d Automatic zero tracking Azn 1d Azn 2d Azn 4d 0 Auto 0 auto 0 Zero setting range after switching on the scales to zero 0 auto 2 0 auto 5 0 auto 10 0 auto 20 0 manl 0 manl 0 Zero setting range the display is set to zero by pressing 0 manl 2 0 manl 4 0 manl 10 0 manl 50 0 manl 100 Pi n Pi n 1 Enter new password...

Page 25: ... out of range Make sure platform empty and check the pan is installed proper Check the load cell connectors Err 9 Unstable Reading Check any air variation vibration RF noise and touching some where Check the load cell and connecters Err 17 Tare out of range Remove the load and restart scale again ol Over range Remove the load Re calibrate Fai l h fai l l Calibration Error Re calibrate Err p Printe...

Page 26: ...C100 Series Technical Manual 23 11 DRAWING ...

Page 27: ...00034 40 kg 14 Star Screw 4 M4x16 15 Load cell lower bracket 1 16 Self thread Screw 4 4x10 17 Insulated Washer 4 8x3 1x1 2t 18 Power Socket 1 19 Power Socket Spacer 1 20 Power Switch 1 899 300077 21 Foot 4 899 300043 22 Key board 1 899 300075 23 Front display overlay 1 24 Level bubble 1 14 7mm 25 Screw 1 M4x35 26 Front Display PBC 1 899 300074 27 Insulated Washer 5 8x3 1x1 2t 28 Star Self thread s...

Page 28: ...ells Connection 4 wire connection to load cells plus shield Maximum cable length 15 Feet Termination DB9 plug on scale Base Specification for Local Scale Load Cell Model No C2X1 Rated Capacity 6 50 kg Rated Out put 2 0 mV V 0 2 mV V Excitation Voltage 20 VDC IP Level IP64 Material Aluminum Alloy Cable Φ 8 2 four core shield Input Resistance 420Ω 30Ω Out put Resistance 350 Ω 5 Ω Temperature Range 1...

Page 29: ...NOTES ...

Page 30: ...C100 Series Technical Manual B TEK Scales 1510 Metric Avenue SW Canton OH 44706 Tel 330 471 8900 Fax 330 471 8909 www B TEK com ...

Reviews: