BMF FREEZER – USER MANUAL
36
3. Temperature Alarm Delay
To set up the time for alarm delay.
1.
Use arrow up or down to change the value.
2.
To save the changed data click on the OK button.
3.
When the high or low temperature level are reached, this alarm delay timer will start
counting.
In this example, where the high temperature alarm level is reached, and the temperature inside
the chamber are not starting to fall under the high temperature alarm level within the 5 minutes,
the High temperature alarm will be activated. You will both see and hear an alarm.