18
5 Alarms
CAUTION
• Upon the occurrence of an alarm situation, attempts must be made to discover the reason for the alarm
and resolve it as quickly as possible. If that is not successful the necessary measures must immediately
be taken so that the stored goods are not damaged.
• An error of the system or a component can lead to a system failure. Depending of the failure type the
error can generate an alarm or a warning
5.1 Messages overview
Code Type Description
E0.1 Alarm Low temperature alarm
E0.2 Alarm High temperature alarm
E0.3 Warning Temperature alarm disabled
E1.1 Alarm Display sensor not connected
E1.2 Alarm Display sensor short circuit
E1.3 Alarm Display sensor reading error
E1.4 Warning Display sensor below reading range
E1.5 Warning Display sensor above reading range
E2.1 Alarm Heating sensor not connected
E2.2 Alarm Heating sensor short circuit
E2.3 Alarm Heating sensor reading error
E2.4 Warning Heating sensor below reading range
E2.5 Warning Heating sensor above reading range
E3.1 Alarm Cooling sensor not connected
E3.2 Alarm Cooling sensor short circuit
E3.3 Alarm Cooling sensor reading error
E3.4 Warning Cooling sensor below reading range
E3.5 Warning Cooling sensor above reading range
E4.1 Alarm Battery not connected
E4.2 Alarm Battery failure
E5.3 Warning Compressor error: motor start error
E5.4 Warning Compressor error: minimum motor speed error
E5.5 Warning Compressor error: thermal cut-off
E9.0 Alarm Communication error
E9.1 Alarm Parameter inconsistency
E9.2 Alarm Flash memory inconsistency
E9.3 Warning System hardware error