86
“OK” after the Low Point(V).
2.
Press Cal. softkey to start calibration. The instrument will output ML Point voltage.
Use the number keys to enter the voltage shown on the DMM. Use the number keys
to enter the voltage shown on the DMM. Press ENTER, and the instrument will show
“OK” after the ML Point(V).
3.
Press Cal. softkey to start calibration. The instrument will output MH Point voltage.
Use the number keys to enter the voltage shown on the DMM. Use the number keys
to enter the voltage shown on the DMM. Press ENTER, and the instrument will show
“OK” after the MH Point(V).
4.
Press Cal. softkey to start calibration. The instrument will output High Point voltage.
Use the number keys to enter the voltage shown on the DMM. Use the number keys
to enter the voltage shown on the DMM. Press ENTER, and the instrument will show
“OK” after the High Point(V).
The calibration date and time will be shown at the bottom of the screen.
15.1.4
AC Current Calibration
Connect a 3 Ω 5,600 watt resistive load to the output of the instrument, and connect a
0.1 mΩ (rated for 30 A) resistive shunt in series to measure the current.
Summary of Contents for 9830B Series
Page 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Page 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Page 28: ...15 6 Menu Tree ...
Page 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Page 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Page 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Page 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...