B+K precision 2165A Instruction Manual Download Page 10

 

10 

 

CONTROLS AND INDICATORS 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

The front panel controls of the instruments are shown in 
Figures 3 through 6 and are explained below. 
 

1.

 

FOCUS.

 Beam sharpness adjustment. 

 

2.

 

INTENS.

 Beam intensity adjustment. 

 

3.

 

POWER

 (Power ON and OFF)

.

 If power is switched to 

ON position, a beam will be visible on the screen after 
approximately 10 sec.. 

 

4.

 

TR (Trace Rotation).

  Despite Mumetal-shielding of 

the CRT, effects of the earth’s magnetic field on the 
horizontal trace position cannot be completely avoided. 
A potentiometer accessible through an opening can be 
used for correction. Slight pincushion distortion is 
unavoidable and cannot be corrected. 

 

5.

 

MARKER  -  ON/OFF switch.

  When the 

MARKER

 

pushbutton is set to the 

OFF

 position, the 

CF

 indicator 

is lit and the display shows the center frequency. When 
the switch is in the 

ON

  position, 

MK

  is lit and  the 

display shows the marker frequency. The marker is 
shown on the screen as a sharp peak.  The marker 
frequency is adjustable by means of the 

MARKER

 

knob and can be aligned with a spectral line. 

 

NOTE 

 

Switch off the marker before taking 
correct amplitude readings. 

 

6.

 

CF/MK (CENTER FREQUENCY/ MARKER) 
indicator.

  The 

CF

  LED is lit when the digital display 

shows the center frequency. The center frequency is 
the frequency which is displayed in the horizontal 
center of the CRT. The 

MK

  LED is lit when the 

Marker pushbutton is in the 

ON

  position. The digital 

display shows the marker frequency in that case. 

 

7.

 

DIGITAL DISPLAY (Display of Center Frequency/ 
Marker Frequency)

  7-segment display with 100 kHz 

resolution. 

 

8.

 

UNCAL.

  Blinking of this LED indicates incorrectly 

displayed amplitude values. This is due to scanwidth 
and filter setting combinations which give uncalibrated 
amplitude readings because the IF-filters have not 
settled. This may occur when the scanned frequency 
range (

SCANWIDTH

) is too large compared to the IF 

bandwidth, and/or the video filter bandwidth. 
Measurements in this case can either be taken without 
a video filter or the scanwidth has to be decreased. 

 

9.

 

CENTER FREQUENCY –  Coarse/Fine.

  Both rotary 

knobs are used for center frequency setting. The center 
frequency is displayed at the horizontal center of the 
screen. 

 

10.

 

BANDWIDTH.

  Selects the IF bandwidth. When the 

switch is engaged, the noise level decreases and the 
selectivity is improved. Spectral lines which are 
relatively close together can be distinguished. As the 
small signal transient response requires a longer time, 
this causes incorrect amplitude values if the scanwidth 
is set at too wide a frequency span. The 

UNCAL

  LED 

will indicate this condition. 

 

11.

 

VIDEO FILTER.

  The video filter may be used to 

reduce noise on the screen. It enables small level 
spectral lines to become visible which normally would 
be within or just above the medium noise level. The 
filter bandwidth is 4 kHz. 

 

12.

 

Y-POS.

  Control for adjusting the vertical beam 

position. 

 

13.

 

INPUT.

 The BNC 50

Ω input of the spectrum analyzer. 

Without input attenuation the maximum permissible 
input voltages of ±25V DC and +l0 dBm AC must not 
be exceeded. With the maximum input attenuation of 
40 dB the maximum input voltage is +20 dBm. 

 

The maximum dynamic range of the instrument is 70 
dB. Higher input voltages exceeding the reference level 
cause signal compression and intermodulation. Those 
effects will lead to erroneous displays. If the input 
level exceeds the reference level, the input level 
attenuation must be increased. 

 

14.

 

ATTN. (ATTENUATOR).

  The Input Attenuator 

consists of four 10 dB attenuators, reducing the signal 
height before entering the 1st mixer. Each attenuator is 
active if the push button is depressed. 

 

The correlation of selected attenuation, reference level 
and baseline level (noise level) is according to the 
following listing: 
 

Attenuation 

Reference level 

Base line 

0 dB 

–27 dBm 

10 mV 

–107 dBm 

10 dB 

–17 dBm 

31.6 mV 

–97 dBm 

20 dB 

–7 dBm 

0.1 V 

–87 dBm 

30 dB 

+3 dBm 

316 mV 

–77 dBm 

40 dB 

+13 dBm 

1 V 

–67 dBm 

 

Summary of Contents for 2165A

Page 1: ...INSTRUCTION MANUAL SPECTRUM ANALYZERS MODELS 2165A 2620A 2625 2630 ...

Page 2: ...some power input circuits such as the on off switch fuses power transformer etc any time the equipment is connected to an ac outlet even if the equipment is turned off 6 Some equipment with a two wire ac power cord including some with polarized power plugs is the hot chassis type This includes most recent television receivers and audio equipment A plastic or wooden cabinet insulates the chassis to...

Page 3: ...3 Instruction Manual for Models 2615A 2620A 2625 2630 SPECTRUM ANALYZERS ...

Page 4: ...AND INDICATORS 10 CALIBRATION 15 Vertical Calibration 15 Horizontal Calibration 15 Page INTRODUCTION TO SPECTRUM ANALYSIS 17 General 17 Types of Spectrum Analyzers 17 Spectrum Analyzer Requirements 18 Frequency Measurements 18 Resolution 18 Sensitivity 19 Video Filtering 19 Spectrum Analyzer Sensitivity 19 Frequency Response 20 Tracking Generators 20 APPENDIX dBm CONVERSION 22 CUSTOMER SUPPORT 23 ...

Page 5: ...ls 2625 and 2630 99 dBm 20 kHz BW Models 2615A and 2620A Distortion 75 dBc 2nd and 3rd harmonic 3rd order intermod 70 dBc two signals 3 MHz apart Sensitivity dB above average noise level Log scale fidelity 2 dB without attn Ref 250 MHz IF gain 10 dB adjustment range Input Input impedance 50Ω Input connector BNC Input attenuator 0 to 40 dB 4 x 10 dB steps Input attenuator accuracy 1 dB 10 dB step M...

Page 6: ...o components or radiation sources The electric field monopole probe has the highest sensitivity of all three probes It can be used to check screening and perform pre compliance testing on a comparative basis The high impedance probe is used to measure directly on the components under test or at the conductive trace of a PC board It has an input capacitance of only 2 pF and supplies virtually no el...

Page 7: ... 1 Safety requirements for elec trical equipment for measurement control and laboratory use The CENELEC regulations EN 61010 1 correspond to this standard It has left the factory in a safe condition This instruction manual contains important information and warn ings which have to be followed by the user to ensure safe operation and to retain the Spectrum Analyzer in a safe condition The case chas...

Page 8: ...ment should be cleaned regularly with a dusting brush Dirt which is difficult to remove on the casing and handle the plastic and aluminum parts can be removed with a moistened cloth 99 water 1 mild detergent Spirit or washing benzene petroleum ether can be used to remove greasy dirt The screen may be cleaned with water or washing benzene but not with spirit alcohol or solvents it must then be wipe...

Page 9: ...rred directly or via a low pass filter to another amplifier This amplifier output is connected to the Y deflection plates of the CRT The X deflection is performed with a ramp generator voltage This voltage can also be superimposed on a dc voltage which allows for the control of 1st LO The spectrum analyzer scans a frequency range depending on the ramp height This span is determined by the scanwidt...

Page 10: ... to the IF bandwidth and or the video filter bandwidth Measurements in this case can either be taken without a video filter or the scanwidth has to be decreased 9 CENTER FREQUENCY Coarse Fine Both rotary knobs are used for center frequency setting The center frequency is displayed at the horizontal center of the screen 10 BANDWIDTH Selects the IF bandwidth When the switch is engaged the noise leve...

Page 11: ...aticule line The lowest horizontal graticule line indicates the baseline The vertical graticule is subdivided in 10 dB steps As previously pointed out the maximum permissible input voltages may not be exceeded This is extremely important because it is possible that the spectrum analyzer will only show a partial spectrum of currently applied signals ...

Page 12: ...lied with excessive levels outside the displayed frequency range leading to the destruction of the input attenuator and or the 1st mixing stage Also refer to INPUT The highest attenuation 4 x 10 dB and the highest usable frequency range highest scanwidth setting should be selected prior to connecting any signal to the input ...

Page 13: ...creased by means of the button and de creased by means of the button Switching is accomplished in 1 2 5 steps from 50 kHz div to 50 MHz div The width of the scan range is displayed in MHz div and refers to each horizontal division on the graticule The center frequency is indicated by the vertical graticule line at middle of the horizontal axis If the center frequency and the scanwidth setting are ...

Page 14: ...on of buttons may be used to achieve the desired attenuation 22 OUTPUT 50Ω BNC output of the tracking generator 23 TRACK GEN The tracking generator is activated when this button is engaged When activated a sine signal can be obtained at the OUTPUT BNC connector at a frequency determined by the spectrum analyzer In ZERO SCAN mode the center frequency appears at the output 24 LEVEL This knob adjusts...

Page 15: ... to the spectrum analyzer input 13 The frequency of this signal should be between 2 MHz and 250 MHz Set the center frequency to the signal frequency A A single spectral line 27 dBm appears on the screen The spectral line maximum is now adjusted with the Y POS control 12 and placed at the top graticule line of the screen All input attenuators switches have to be released B Next the generator signal...

Page 16: ...frequency to 250 MHz Adjust the peak of the 250 MHz spectral line to the horizontal screen center using the X POS control 16 D Set the generator frequency to 50 MHz If the 50 MHz spectral line is not on the 2nd graticule line from left it should be aligned using the X AMPL control 17 Then the calibration as described under C should be verified and corrected if necessary The calibrations C and D sh...

Page 17: ...e spectrum analyzer can be used to measure long and short term stability Parameters such as noise sidebands on an oscillator residual FM of a source and frequency drift during warm up can be measured using the spectrum analyzers calibrated scans The swept frequency responses of a filter or amplifier are examples of swept frequency measurements possible with a spectrum analyzer These measurements a...

Page 18: ...ty is usually characterized as either short term or long term Residual FM is a measure of the short term stability which is usually specified in Hz peak to peak Short term stability is also characterized by noise sidebands which are a measure of the analyzers spectral purity Noise sidebands are specified in terms of dB down and Hz away from a carrier in a specific bandwidth Long term stability is ...

Page 19: ... power average noise power The analyzer always measures signal plus noise Therefore when the input signal is equal to the internal noise level the signal will appear 3 dB above the noise When the signal power is added to the average noise power the power level on the CRT is doubled increased by 3 dB because the signal power average noise power The maximum input level to the spectrum analyzer is th...

Page 20: ...tor is the spectrum analyzer lst LO The spectrum analyzer tracking generator system is used in two configurations open loop and closed loop In the open loop configuration unknown external signals are connected to the spectrum analyzer input and the tracking generator output is connected to a counter This configuration is used for making selective and sensitive precise measurement of frequency by t...

Page 21: ...21 Introduction to Spectrum Analysis Fig 7 Block Diagram ...

Page 22: ... 15 8 μV 7 40 100 100 mV 46 1 61 1122 μV 84 37 23 14 1 μV 8 39 99 89 1 mV 47 0 60 1000 μV 85 38 22 12 6 μV 9 38 98 79 4 mV 48 1 59 891 μV 86 39 21 11 2 μV 10 37 97 70 7 mV 49 2 58 794 μV 87 40 20 10 0 μV 11 36 96 63 1 mV 50 3 57 707 μV 88 41 19 8 91 μV 12 35 95 56 2 mV 51 4 56 631 μV 89 42 18 7 94 μV 13 34 94 50 1 mV 52 5 55 562 μV 90 43 17 7 07 μV 14 33 93 44 7 mV 53 6 54 501 μV 91 44 16 6 31 μV ...

Page 23: ...or this service We maintain a network of B K PRECISION authorized service agencies for this purpose To use this service even if the instrument is no longer under warranty follow the instructions given in the WARRANTY SERVICE INSTRUCTIONS portion of this manual There is a nominal charge for instruments out of warranty Technical advice on the use of your instrument Technical advice on special applic...

Page 24: ...bes connectors and accessories that you are using with the device Customers not on an open account must include payment in the form of a money order or credit card For the most current repair charges please refer to the service and support section on our website Return all merchandise to B K Precision Corp with pre paid shipping The flat rate repair charge for Non Warranty Service does not include...

Page 25: ...form on our website www bkprecision com within fifteen 15 days of purchase Exclusions This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs The warranty is void if the serial number is altered defaced or removed B K Precision Corp shall not be liable for any consequential damages including without limitation damages result...

Page 26: ...26 NOTES ...

Page 27: ...27 ...

Page 28: ...28 22820 Savi Ranch Parkway Yorba Linda CA 92887 v110409 Printed in China ...

Reviews: