- 43 -
With the right arrow key, move the cursor (underline) to the next position.
The minutes are displayed. Set the minutes with the arrow up/down keys.
With the cursor over the right arrow key, go to the next position ”Year“ and set it with the
up/down keys.
Afterwards, set month and day as described above.
The changes are immediately active.
4.2.6. Language
The desired language for menus and warning messages can be configured under the menu
item ”Language”.
With the right arrow key, activate ”Main Menu”, highlight the ”Setup” menu with the
down key and open with the right arrow key. With the up or down key, highlight the menu
”Language” and open with the right arrow key. Now, you have the option of highlighting the
desired language with the up or down key and activating your selection with the right arrow
key (
appears on the right).
When switching to
an
other language than English, the English terms for Setup and Language
menus are always additionally displayed behind the terms for the selected language to
enable the user to reset the language when he unintentionally changed it.
4.2.7. Options
This menu page offers the following options:
1. Factory standard restores factory settings.
2. Electrical charge nC – additional display of the amount of electricity to be given off in
nanocoulomb.
3. Auto current adjustment – the current is adjusted automatically when the stimulus du-
ration is changed.
4. Add. stimulus duration – additional stimulus duration parameters.
5. SENSe - SENSe option can be enabled / disabled
Use the right arrow key to activate (
appears on the right) or deactivate these options
(
not visible).