3.2 Battery Module
Designation
Ord. No.
Battery compartment cover PSP, cpl. . . . . . . . . . . . . . 3452 0872
Battery module SP (NiMH)
without fixing pin. . . . . . . . . . . . . . . . . . . . . . . . . 3452 0856
with fixing pin. . . . . . . . . . . . . . . . . . . . . . . . . . . 3452 0856A
Battery module SP (Li-Ion)
with WLAN and fixing pin . . . . . . . . . . . . . . . . . . . 8713182A
Note
Users have been informed as follows: If a Perfusor® Space pump
used on battery only and showing full battery charge stops during
syringe change and alarms, users should remove the pump from
use and return it to the Medical Engineering Department for
replacement of the battery.
In this case the battery has to be replaced regardless of its age or
any further test.
Disassembly
Note
Move the drive head to the extended end position before starting
disassembly
(see “General“
➨
pg. 3 - 1)
.
1. Press the lock (
Fig.: 3 - 8 / Item 2
) on the battery compart-
ment cover (
Fig.: 3 - 8 / Item 1
) and remove the battery
compartment cover.
Note
The battery compartment cover of more recent devices is shorter
and can only be unlocked using a pointed object.