Procedural Instructions on the TSC
8
Infusomat® fmS 2.3 gb
8 - 3
Functional Inspection
Switch on Unit
1.
Switch on unit and check the following details:
-
Self-test
-
Audible alarm
One short and afterwards two short tones
-
Status displays
LEDs (red, green) light up for a short moment
-
LCD lighting
-
Display on the LC display
All symbols and display pixels flash for a short moment
. Then „000.0ml/h“ is displayed
Operation
1.
Open door.
Pump cover must automatically open when the unit door is
opened.
2.
Insert the Original Infusomat Line (OIL). If the unit is designed
for an Infusomat® Space Line, then insert an Infusomat®
Space Line.
3.
Close door.
The door must lock smoothly and close correctly at the top
and bottom.
4.
Carry out infusion with intermediate bolus and press all but-
tons at least once.
Infusion and bolus are performed and all buttons trigger the
function desired.
5.
Plug MFC service connector on the MFC socket.
6.
Trigger a staff call (e.g. open pump cover during operation).
The red LED in the MFC service connector lights up.
- if ”dynamic” is set, 1 sec.
- if ”static” is set, until the alarm is acknowledged.
Check the MFC staff call line if installed.
7.
Press the alarm button.
The current alarm is suppressed for 2 minutes.
Fig.: 8 - 2
Summary of Contents for Infusomat fmS
Page 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10...
Page 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes...
Page 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes...
Page 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes...
Page 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes...
Page 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes...
Page 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes...
Page 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes...
Page 84: ...Index 11 11 2 Infusomat fmS 2 3 gb...
Page 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes...