D I S A S S E M B L Y

Deactivate ratchet on ratchet handle 

Flip down red toggle switch.

N.B.: for PO959R only

I

Disconnect shaft and handle

Pull rotation knob back until it locks in place. A red
ring becomes visible.

Important:

Do not hold on to the movable handle component.

II

Assemble shaft II

Slide the outer insulation tube as far as it will go
over the inner metal tube.

II

Remove shaft from handle

Remove the shaft by pulling the 
instrument shaft out of the handle. 

III

Disassemble shaft I

Hold the jaw insert at the tip and remove the outer
insulation tube.

IV

Prepare handle

If the red ring is not visible, pull the rotation knob
back. Deactivate the ratchet by flipping down the 
red toggle switch. 

N.B.: for PO959R only

Connect shaft and handle 

Hold the shaft firmly by its tip and the handle behind
the rotation knob. Push the shaft into the handle until
it locks in place and the red ring disappears.

Important:

Do not hold on to the movable handle component.

Disassemble shaft II

Hold the jaw insert at the tip and remove 
the inner metal tube.

V

A S S E M B L Y

C A R E

R E P A I R S

Reprocess the product immediately after use

The product can be sterilized either in disassembled or in assembled condition

For advice on cleaning, disinfection and sterilization, please follow the instructions for use

For the latest information on reprocessing, please also refer to the Aesculap Extranet via www.aesculap-extra.net

For accessories and spare parts, please refer to Brochure C766 or visit www.endoscopy-catalog.com

P R O C E S S I N G   &   A C C E S S O R I E S

Lubricate the defined places on the
instruments (instruments jaw and
red toggle switch) with Aesculap
Sterilit Oil after each cleaning cycle.

Aesculap Endoscopic Technology

V

Function test

Test the assembly of the instrument by opening and
closing the handle to ensure the jaw is opening and
closing.

In order to ensure the preservation of full guarantee and 
warranty rights, only persons authorized by Aesculap may 
carry out repairs.

Please send products for repair to: Aesculap Technical Service,
Am Aesculap-Platz, 78532 Tuttlingen, Germany.

Additional service addresses may be obtained from the same
address or via [email protected]

I

Assemble shaft I

Slide the metal inner tube onto the jaw insert 
with its four spring tongues towards the working
end until the tongues snap in place.

jaw insert

inner metal tube

outer insulation tube

IV

���

toggle switch

���

rotation knob

���

movable handle component

���

���

���

III