![B. Braun Aesculap Acculan 4 GA346 Instructions For Use/Technical Description Download Page 5](http://html1.mh-extra.com/html/b-braun/aesculap-acculan-4-ga346/aesculap-acculan-4-ga346_instructions-for-use-technical-description_3108360005.webp)
5
7.5
Wipe disinfection for electrical devices without sterilization
RT: Room
temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
7.6
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
►
Set aside the product if it is damaged.
8.
Maintenance
The display "Battery replacement recommended" on the Acculan 4 charger
GA320 (see TA014534/TA014535) indicates when the battery should be
replaced due to reduced capacity.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting
►
Replace product: have defective products repaired by Aesculap Techni-
cal Service, see Technical Service.
CAUTION
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
►
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disin-
fecting agent.
►
Do not introduce any fluid onto the electrical
contacts of the battery.
Phase
Step
D
[°C/°F]
t
[min]
Conc.
[%]
Water quality
Chemistry
I
Wipe disinfection
RT
≥
1
-
-
Meliseptol Wipes sensitive
Issue
Detection
Cause
Remedy
Battery not working
No signal tone when inserting the
battery into the machine
Battery defective
Replace product.
Battery quickly empty
Symbol "Battery replacement rec-
ommended" illuminates at the end
of the charging process on the char-
ger GA320
Battery cells at end of life
Have battery cells replaced by
Aesculap Technical Service.