Intended applications

 

!

The intended applications cover contact less measurement of temperature from -40 °C to +816 °C. For power supply, only two 1.5 V bat-
teries of type Mignon AA, IEC LR6 or equivalent are to be used. The operation is allowed only under dry environment conditions, contact 
with moisture should be essentially avoided. A measurement under unfavourable environment conditions is not allowed. Unfavourable 
environment conditions are: Dust and combustible gases, steam or solvents, thunderstorms as well as thunderstorm conditions like strong 

electrostatic fields etc.
Any other application, as described before, can lead to damage of this product. In addition, dangers like short-circuit, fire, etc. are also 
involved. The product should not be opened, modified or re-engineered!

Steps to eliminate measuring errors

Please take care of the following instructions while using the instrument: 
• 

Avoid using the device in the vicinity of electrical welding sets, induction heaters, transmission equipment, frequency converters, and 

other strong electromagnetic fields.

• 

After any abrupt change in temperature, the device should be left out for 15 minutes before using so that it can stabilise itself to the 
new environment temperature.

• 

Do not expose the equipment to high temperatures for a long time.

• 

Avoid dusty and humid environment conditions. After usage, store the device in the storage bag to avoid contamination of the lens.
Consider technical facts of the measuring procedure to obtain exact results of measurement, particularly the relationship of measuring 
spot size with respect to distance and the information for adjustment of emission value.

4

Working principle

Every body and liquid above the temperature of absolute zero degree Kelvin (-273.15 °C) emits infrared radiations from its surface. The 
emitted energy has a fixed mathematical relationship with the surface temperature. This energy is measured by our infrared thermometer, 

electronically processed and shown as temperature value on the display.

4.1

Measuring spot size

In order to achieve exact results from measurements, the measuring object must be bigger than the measuring spot of the infrared ther-
mometer. The measured temperature is the average temperature of the object surface. Smaller the object, shorter should be its distance 
from the thermometer.
The exact measuring spot size can be decided from the diagram shown below.  This diagram is also shown on the side of the device.

5

Setting-up and operation

5.1

Control elements

B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen 
Fon +49 771 83160 | Fax +49 771 8316-50 | [email protected] | bb-sensors.com

5 / 12

Summary of Contents for IR 8895

Page 1: ...nical changes reserved Infrared thermometer IR 8895 0141 0316 190 30 11 2021 B B Thermo Technik GmbH Heinrich Hertz Straße 4 D 78166 Donaueschingen Fon 49 771 83160 Fax 49 771 8316 50 info bb sensors com bb sensors com 1 12 ...

Page 2: ... 5 Setting up and operation 5 5 1 Control elements 5 5 2 Inserting batteries 6 5 3 Laser beam activation de activation 6 5 4 Temperature measurement 7 5 5 Changeover from C to F 7 5 6 Background illumination 8 5 7 Emission setting 8 6 Maintenance and cleaning 8 6 1 Cleaning 8 6 2 Replacing batteries 9 6 3 Disposal of used batteries 9 6 4 Scrapping and disposal 9 7 Technical data 10 8 Guarantee 10 ...

Page 3: ...nable you to carry out exact temperature measurements even for highly demanding industrial applications Of course the instrument also complies with the current safety and environmental protection regulations and is conforming to RoHS This instruction manual made available to you shall help in adjustment and operation of the instrument The information is important to achieve optimum results and als...

Page 4: ...ation of the device the safety instructions warnings and instructions provided in chapter on Application purpose shall be necessarily followed 2 1 General safety instructions The guarantee claim becomes void in case of damages caused due to non compliance to the instructions We take no liability for the consequential damages which result from such negligence This appliance is with CE conformance a...

Page 5: ...t can stabilise itself to the new environment temperature Do not expose the equipment to high temperatures for a long time Avoid dusty and humid environment conditions After usage store the device in the storage bag to avoid contamination of the lens Consider technical facts of the measuring procedure to obtain exact results of measurement particularly the relationship of measuring spot size with ...

Page 6: ... the instrument for the first time insert two new 1 5 V Batteries type Mignon AA Instructions on inserting batteries have been described in detail in the section Maintenance and cleaning 5 3 Laser beam activation de activation For a confidence of having done an exact measurement the infrared thermometer 8895 is provided with a red point laser This laser beam approximately shows the centre of the m...

Page 7: ...he hottest point on the object to be measured the infrared thermometer 8895 can be pointed towards any point lying outside the measuring range and then with the trigger in pressed condition the desired range is to be scanned with zigzag movements till the hottest point is located Make sure that the measured object is not smaller than the measuring spot size for a given distance 5 5 Changeover from...

Page 8: ...t finish paint or black adhesive film The device can not measure through transparent surfaces like for example glass or transparent film Instead the surface temperature of the glass or the transparent film gets measured For entering the respective emission factor the device is to be first activated by pressing the trigger Now press the Mode Save key C The emission symbol 6 begins to blink Then ent...

Page 9: ...not been used for long because even leak proof batteries can cause corrosion Chemicals which can da mage the device or cause health hazards can leak through the batteries 6 3 Disposal of used batteries The disposal of batteries together with house garbage is not allowed Batteries containing harmful materials are marked with the symbol as shown on left which refer to the prohibition of dispoal with...

Page 10: ...aged or removed Apart from the guarantee service if any essential repairs are required to be carried out the service is free However further services and also postage and packing expenses are chargeable Compensation demands on the basis of claim for liability or damages during the guarantee period are excluded and these are in general not legally covered 9 Repair and calibration services During th...

Page 11: ...tainless steel assorted 0 2 0 6 10 2 Brass Material Emission value Brass polished 0 1 Brass raw surface 0 2 Brass oxidised 0 6 10 3 Aluminium Material Emission value Aluminium polished 0 1 Aluminium severely oxidised 0 25 Aluminium oxide at 260 C 0 6 Aluminium oxide at 800 C 0 3 Aluminium alloy assorted 0 1 0 25 10 4 Copper Material Emission value Copper polished 0 05 Copper oxidised 0 8 Copper gl...

Page 12: ...d 0 55 Zirconium silicate up to 500 C 0 85 Zirconium silicate at 850 C 0 6 Quartz raw 0 9 Coal Graphite 0 75 Material Emission value Coal Carbon 0 95 Timber assorted 0 8 0 9 Enamel all colours 0 9 Oil paint all colours 0 95 Lacquers 0 9 Black colour matt finish 0 95 0 98 Aluminium paint 0 5 Water 0 98 Rubber finished 0 9 Rubber raw 0 98 Plastic different solids 0 8 0 95 Plastic sheets 0 05 mm thic...

Reviews: