Chapter 4. OPERATION
Displaying the total flow rate
This function displays the total flow rate since the device was installed.
Reset cannot be performed by the same reset operation used for inte-
grated flow rate.
(1) Press the key until the integrated flow rate is displayed.
The m
3
indicator lights.
(2) Hold the
key for one second or more.
The m
3
indicator goes out, and the total flow rate is displayed for
5 seconds.
The total flow, like the integrated flow rate, is displayed in the
divided upper four digits and lower four digits, in total 8 digits
capable of displaying the maximum m
3
.
After that, the m
3
indicator automatically lights, and the display
returns to the integrated flow rate display.
(3) Press the key again within five seconds while it is displayed.
The first four digit display.
You can also alternately display the first four digits and the last
four digits by using the
key repeatedly.
Handling Precautions
•
The total flow rate indication returns to 0 after
99999999
is
exceeded.
•
Integrated flow rate and total flow rate values are held in memory
even if the power is turned OFF.
On models that display values down to two digits past the decimal
point, data is written into memory when the ones digit changes or
one hour after the previous writing.
On models that display values down to one digit past the decimal
point, data is written into memory when the tens digit changes or
one hour after the previous writing.
Integrated values that have not been written to memory are dis-
carded when the power is turned OFF.