2-26
Accessory parts for installation
Introduction
The following parts are necessary for the installation of the detector:
Gaskets: Provides gaskets by customer when using grounding rings made of SUS material.
Supplied gaskets as standard accessory, when using grounding rings made of
other material.
Gaskets
Gaskets are supplied with the grounding ring, except when it is made of SUS material.
Provides the gaskets by customer when you use a grounding ring made of SUS material.
We recommend a non-rubber gasket material such as joint sheet or PTFE.
For the internal diameters of the gaskets, refer to Table. 2-6.
Note) A gasket with small internal diameter may generate turbrent flow and affect, resulting in
inaccurate measurements.
A gasket diameter with two large internal may cause leakage. Also, if there are any
solids in the fluid to be measured, these may build up between the gasket and the flange,
resulting in inaccurate measurements.
Table. 2-6. Recommended internal diameters of gaskets
Body diameter
Internal diameter (mm)
2.5 mm
11±1
5 mm
11±1
10 mm
11±1
15 mm (1/2 inch)
16±1
25 mm (1 inch)
25±1
40 mm (1-1/2 inch)
40±1
50 mm (2 inches)
51±1
65 mm (2-1/2 inches)
64±1
80 mm (3 inches)
76±1
100 mm (4 inches)
101±1
125 mm (5 inches)
124±1
150 mm (6 inches)
148±1
200 mm (8 inches)
196±1
Summary of Contents for MagneW Two-wire PLUS+ MTG15A
Page 8: ...vi...
Page 22: ...1 14...
Page 56: ...2 34...
Page 60: ...3 4 Fig 3 2...
Page 66: ...3 10...
Page 70: ...4 4...
Page 144: ...5 74...
Page 226: ...6 82...
Page 268: ...7 42...
Page 274: ...8 6...
Page 275: ......
Page 278: ......
Page 280: ......