Azbil Corporation
MagneW PLUS+ Electromagnetic Flowmeter Open channel Flowmeter Detector
15
3-4 : How to use the Dummy detector
The measurable flow rate can be doubled, tripled, or quadrupled by installing 1,2,3
dummy detectors, respectively. A detector must be combined with dummy detectors
of the similar shape. If the flow rate is to be increased in future, the measurable
range can be expanded easily by installing a model with an elbow flange. (with a
cover).
[1 flowmeter]
1 flowmeter
1 dummy
1 flowmeter
2 dummies
Q=
q
Q=
2q
Q=
3q
q
q
q
q
q
q
Detector
Dummy
Dummy
Detector
Figure 9
Measuring a large flow rate using dummy detectors
Flow rate measurement in trunk sewerage
(400mm: one detector and three dummies)
Flow rate measurement in sewage treatment
plant
(600mm: one detector and nine dummies)
Figure 10
Examples of setting Dummy detector