Azbil CY8100C1000 Specifications & Instructions Download Page 7

AB-6256 

Maintenance 

Periodic inspection 

Since the measurement gases flow into the inside of the housing, annual cleaning of circuit boards inside the housing is 
recommended. 

 

Periodic calibration 

The output drift (a maximum of 150 ppm/year) occurs even if this product is in continuously energized state.  Be sure to conduct 
a CO

2

 zero gas calibration every year. 

IMPORTANT: 
 As 

mentioned 

in 

Installation

 section, leaving the main unit non-energized for a long time may cause drift in 

Table 1.  In such a case, CO

2

 zero gas calibration of the main unit must be carried out after the supply of power. 

Note that the output needs about ten days to stabilize for the main unit continuously non-energized for a long
time.  Please calibrate the main unit when it has been continuously energized for 10 days. 

 

The zero calibration service bag (Part No. 83104511-001) needs to be ordered for the main unit calibration. 
Refer to the following procedure and perform CO

2

 zero gas calibration. (See Figs. 9 and 10) 

1)  Remove the front cover from the main unit. 

2)  Remove the adaptor attached to the rear side of the front cover, and attach the adaptor to the CO

2

 zero gas inlet on the main 

unit.  Connect the zero calibration service bag with the main unit CO

2

 zero gas inlet, referring to Specifications/Instructions 

of the zero calibration service bag (AB-5803).  Two inlets are provided on the main unit.  Use either one. 

3)  Turn on the zero calibration service bag to start supplying CO

2

 zero gas to the main unit. 

4)  Keep supplying the CO

2

 zero gas until CO

2

 concentration is stabilized (approximately for five minutes). 

5) When 

CO

2

 concentration is stabilized, calibrate the main unit referring to Fig. 10. 

6)  Yellow LED lights up to indicate when calibration is completed successfully. Then, the LCD returns to display the CO

concentration. 

7)  Remove the adaptor from the CO

2

 zero gas inlet and attach it to the rear side of the front cover. 

8)  Put the front cover back to the main unit. 

Refer to AB-5803 for handling the zero calibration service bag (Part No. 83104511-001). 

 

Layout of CO

2

 zero gas inlets, buttons, and LED 

       Operation flow of CO

2

 zero gas calibration and error code check 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CO

2

 zero gas inlets 

(Use either one for CO

2

 zero gas calibration.) 

+ button 
 
- button 
 
Menu button

Yellow LED
 
Green LED 

Figure 9.  Layout of CO

2

 zero gas inlets, buttons, and LED

on the main unit (without the front cover) 

(Menu)

(Menu)

(-) 

(Menu) 

CO

2

 concentration display [XXXX ppm]

Error code [EXXX] 

Error code reset

Zero gas calibration [CAL] 

Confirmation of zero gas  

calibration excecution [SURE]

 

Execution of  

zero gas calibration 

(+) 

(-)

(+)

(+)

(Menu)

(-) 

(Menu)

(-) 

(Menu): Press Menu button. 
(+): Press + button. 
(-): Press - button. 

Notes: 

  Indication on LCD returns to CO

2

 concentration if no button is 

pressed for 10 seconds. 

  Let go of the main unit when LED turns on. 

  Wait until the LED turns off, then go to the next operation. 

[   ]: Indication on the LCD. 
      : CO

2

 zero gas calibration

Figure 10.  Operation flow of  

CO

2

 zero gas calibration and error code check. 

Summary of Contents for CY8100C1000

Page 1: ...cylinder Specifications 1 2 Item Specification Measuring range 0 to 2000 ppm CO2 Concentration Operating principle Non dispersive infrared absorption method NDIR Measuring accuracy 50 ppm 5 of reading...

Page 2: ...damages This product must be operated within its rated operating ranges specified in this manual Failure to comply will cause equipment damage All wiring must comply with local codes of indoor wiring...

Page 3: ...mensions of the main unit with sampling probe mm Assembly kit Sampling probe Cable assembly Figure 2 Dimensions of the assembly kit mm 4P connector shows the mounting orientation 4 4 25 28 28 Mounting...

Page 4: ...er is supplied Assembly The assembly kit sampling probe and cable assembly and the main unit must be ordered and mounted separately to minimize the drift See the above Precautions for installation sec...

Page 5: ...robe by snap in bayonet mount Refer to the following procedure 1 Engage the notches of the main unit with the locking knobs of the sampling probe then lightly press the main unit 2 Turn the main unit...

Page 6: ...it 10 cm away from the main unit Check that the reading on the LCD and the output ascend quickly LCD indication On the LCD CO2 concentration is continuously displayed The display is changed a when a b...

Page 7: ...main unit Use either one 3 Turn on the zero calibration service bag to start supplying CO2 zero gas to the main unit 4 Keep supplying the CO2 zero gas until CO2 concentration is stabilized approximat...

Page 8: ...en the power is turned on or during the normal operation is detected and indicated with the yellow LED and the too icon on the LCD as shown in Fig 11 Figure 11 Tool icon on the LCD If error code excep...

Reviews: