Azbil ATT082 Brief User Manual Download Page 30

Summary of Contents for ATT082

Page 1: ...CM2 ATT082 2002 ATT082 Temperature transmitter Brief User s Manual...

Page 2: ...for its customer In no event shall Azbil Corporation be liable to anyone for any indirect special or consequential damages This information and specifications in this document are subject to change w...

Page 3: ...als 7 3 4 CE mark Declaration of Conformity 8 3 5 HART Protocol Certification 8 3 6 Korea Certification Mark 8 Chapter 4 Installation instructions 9 4 1 Incoming acceptance transport storage 9 4 1 1 I...

Page 4: ...elements 17 6 1 3 Local operation 18 6 2 Configuration of transmitter and HART protocol 18 Chapter 7 Commissioning 19 7 1 Post installation check 19 7 2 Switching on the transmitter 19 7 3 Enabling co...

Page 5: ...ameplate You can use this Ex documentation if the two numbers i e in the Ex documentation and on the nameplate are identical 1 2 Safety Precautions for Use For safe use of the product the following sy...

Page 6: ...stand on this unit Doing so may damage the device or cause injury Bumping the glass of the display with a tool may cause damage or injury Be careful Install the device correctly Incorrect or incomplet...

Page 7: ...working on the vent check its direction so that people do not come into contact with vented gas There is a danger of burns or other physical harm When the device is being used in an explosion proof a...

Page 8: ...4 Chapter 1 Important document information...

Page 9: ...resistance and voltagetransmitters The head transmitter version of the device is intended for mounting in a flat faceterminal head as per DIN EN 50446 The manufacturer is not liable for damage caused...

Page 10: ...6 Chapter 2 Basic safety instructions...

Page 11: ...itter example Ex version 1 Power supply current consumption and extended order code 2 Serial number device revision and firmware version 3 Approvals with symbols 4 2 lines for the TAG name 5 Approval...

Page 12: ...es The manufacturer confirms that the device is compliant with the relevant guidelines by applying the CE mark 3 5 HART Protocol Certification The temperature transmitter is registered by HART Communi...

Page 13: ...mitter version with screw terminals Dimensions in mm in A Spring travel L 5 mm not for US M4 securing screws B Mounting elements for attachable measured value display C Interface for contacting measur...

Page 14: ...uide the mounting screws 2 through the lateral bores of the head transmitter 3 3 Screw the head transmitter to the field housing 4 After wiring close the field housing cover 1 4 4 Post installation ch...

Page 15: ...the terminal head or the field housing 2 Feed the cables through the opening in the cable gland 3 Connect the cables as shown in 5 1 4 Retighten the cable gland and close the housing cover In order to...

Page 16: ...ments considerably The sensors must remain galvanically isolated from one another by con necting each sensor separately to a transmitter The transmitter provides sufficient galvanic isolation 2 kV AC...

Page 17: ...unication Observe grounding concept of the plant Please also observe the general procedure on the page 10 1 2 1 2 3 4 Figure 5 2 Connecting the signal cables and power supply 1 Terminals for HART prot...

Page 18: ...f disturbance variables as per NAMUR NE21 is thus guaranteed 1 2 3 4 Figure 5 3 Shielding and grounding the signal cable at one end with HART communication 1 Optional grounding of the field device iso...

Page 19: ...11 to 42 V DC SIL mode U 11 to 32 V DC for the head transmitter or U 12 to 32 V DC for the DIN rail transmitter Do the cables have adequate strain relief Are the power supply and signal cables correc...

Page 20: ...16 Chapter 5 Wiring...

Page 21: ...PV I e g S1 for a measured value from channel 1 or DT for the device temperature 6 Configuration locked symbol The configuration locked symbol appears when configuration is locked via the hardware 7 S...

Page 22: ...display onto the head transmitter in the correct position The head transmitter accepts the settings within one second 5 Secure the cover back onto the terminal head or field housing Switching write p...

Page 23: ...stic events and the corresponding troubleshooting instructions can be found in the User s Manual The device is operational after approx 30 seconds and the plug in display after approx 33 seconds in no...

Page 24: ...20 Chapter 7 Commissioning...

Page 25: ...rinsically safe systems for Hazardous classified locations and the National Electrical Code ANSI NFPA 70 FM Approved Associated Apparatus must meet the following parameters Uo Ui Io Ii Po Pi Ca Ci Cca...

Page 26: ...ed for appropriate area classification Class II use a dust tight seal Keep tight when circuits alive NONINCENDIVE NI Class I Div 2 Groups ABCD Depending on location install per National Electrical Cod...

Page 27: ......

Page 28: ...onsibility taking the following matters into consideration 1 Regulations and standards or laws that your Equipment is to comply with 2 Examples of application described in any documents provided by Az...

Page 29: ...u are required to renew any Azbil Corporation s products every 5 to 10 years unless otherwise specified in specifications or instruction manuals System products field instruments sensors such as press...

Page 30: ......

Page 31: ...Document Number CM2 ATT082 2002 Document Name ATT082 Temperature transmitter Brief User s Manual Date 1st edition May 2015 2nd edition July 2016 Issued Edited by Azbil Corporation...

Page 32: ......

Reviews: