GB
When finished, press [ TUNE ] key for 2 seconds or leave the unit for
60 seconds to return to normal mode.
Pressing [ MODE / ALERT ] key for 2 seconds during temperature and
humidity calibration mode will reset the value to the actual or
uncalibrated measurement.
Reset the temperature and humidity values for each channel individ-
ually. There is no batch reset.
Calibration mode sequence: channel> temperature > humidity
In temperature calibration mode, press [ CHANNEL / + ] or
[ MEM / - ] key to modify 0.1 °C or °F of the value. Long press to
quick-adjust the value.
3.
4.
5.
6.
7.
In humidity calibration mode, press [ CHANNEL / + ] or [ MEM
/ - ] key to modify 1% of the value. Long press to quick-adjust the
value.
8.
If the current reading of the selected channel displays the flashing “
--”, “ Hi” or “ Lo”, it will be impossible to adjust. Pressing [ CHANNEL
/ + ] or [ MEM / - ] key will not change the value.
If re-enter temperature and humidity calibration mode, the adjusted
value will reset to the original value.
9.
10.
Temperature calibration
Humidity calibration