ACTIVATING THE TRIAL STIMULATOR
The following section describes the process for activating the TS for delivering stimulation. This should be performed when an
Axonics lead has already been implanted and a Clinician Programmer is available.
1. Press the button on the back of the TS (
Figure 2
). The green light next to the button will start to flash. The green light will flash
for 90 seconds.
a. If you connect a CP to the TS before 90 seconds expires, the TS will remain active and can be programmed. After programming
is complete and the TS is disconnected from the CP, the green light will not be on. However, the TS will remain active.
b. If a CP does not connect to the TS, the TS will return to hibernate mode. It can be turned on again by pressing the button on the
back of the TS.
Notes:
• If a red light flashes there is an error with the TS. Connect the CP for detailed error information. Refer to the CP manual for detailed
troubleshooting information.
• If there is no flashing light when you press the button, try pressing the button again. Confirm you are pressing the button as shown
in Figure 2. If there is no light after multiple attempts at pressing the button, do not use this TS. Activate a new TS. If the issue
persists with a new TS, contact Axonics.
Figure 2:
Activate the TS by pressing the button on the back of the stimulator