Steps during calibration with external weight:
Press
Menu
key, choose
Calibration
option with the cursor and press
ENTER
.
Check if
External load
value matches
the value of external weight used for
calibration. If not, choose
External
load
option and enter correct value.
Choose
External calibration
option
with the cursor and press
ENTER
.
Wait until taring process is finished.
Put calibration weight matching
displayed value on.
Take off the calibration weight.
Wait until internal calibration is
finished.
The balance is ready to work.
USER FUNCTIONS&OPT.
Autotaring
PCS counting
. . .
Calibration
. . .
USER FUNCTIONS&OPT. \ CALIBRATION
1. External calibration
2. External load
: 200g
3. Time of calibration
: 2h
4. Temp. of calibration : 2.0
o
C
5. Report printout
6. Exit
Max ... Min ... e= ... d= ...
0.000
g
0%
100%
CALIBRATION
External calibration:
taring ...
CALIBRATION
External calibration:
Put external load _200.000 g
CALIBRATION
External calibration:
take off ext. load
CALIBRATION
Internal calibration:
please wait ...