![Axis ACA1000G User Manual Download Page 56](http://html.mh-extra.com/html/axis/aca1000g/aca1000g_user-manual_3038945056.webp)
56
USER MANUAL
____________________________________________________________________________________________________________________
Applications option:
- Activate
–
density measurement,
-
Preview
–
register check,
-
Type of material
–
solid or liquid,
-
Type of liquid
–
water, ethanol or other
(inscribe density here),
-
Shortcut
–
set fast access key for login user (if
you have USB_A interface and PC keyboard
connected).
Actions order for solid material:
After choosing material type, liquid type or
density, after choosing
Activate
tare the balance
using
T
key
.
Put material on upper pan (measurement in air)
and press
ENTER
.
Put material on lower pan (measurement in
liquid) and press
ENTER
.
Results will display and available options:
-printout,
CLR - exit to summing.
In order to finish work with the app, choose the
app and
Deactivation
.
APPLICATIONS
...
Density
...
Exit
DENSITY
1. Activate
2. Preview
3. Type of material:
<solid>
<liquid>
4. Type of liquid: <Water><Ethanol><Other>
5. Liquid type:
..... g/cm3
6. Shortcut:
<-><F1><F2>...<F5>
7. Exit
Max ... Min ... e= ... d= .. II
AUT DEN
T
0.0000
g
0%
100%
Max ... Min ... e= ... d= .. II
AUT DEN
ENTER
10.0900
g
0%
100%
Max ... Min ... e= ... d= .. II
AUT DEN
ENTER
8.0910
g
0%
100%
DENSITY/PREVIEW
Mass in the air : 10.090g
Mass in the liquid : 8.0910g
Mass density
: 5.0370g/cm3
Density with comp.: 5.0360g/cm3
- print CLR - exit