background image

3. General Maintenance

3.1.

Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.

Take out the batteries if the meter won’t be used for a long time.

Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.

Dirt or moisture in the terminals can affect readings.

Clean the terminals as follows:

Turn the Meter off and remove all test leads.

Shake out any dirt that may be in terminals.

Soak a new swab with alcohol. Clean each terminal with the swab.

3.2. Replacing the Battery

The Meter is powered by two AA batteries (IEC LR6).
Warning
To avoid electrical shock or personal injury:

Remove test leads from the meter before opening the battery door.

Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.

Close and latch the battery door before using the meter.

Note

The new and old Batteries can not be mixed.

Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.

Make sure the battery’s odes are in accordance with the symbols illustrated in battery pool when replacing them.

Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.

Take out the batteries if the meter won’t be used for a long time.

Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.

Dispose the old batteries in accordance with the local law

Replace the batteries as follows.

Turn the rotary switch to OFF and remove the test leads from the terminals.

Take off the protector of the meter , remove the battery door by using a standard-blade screwdriver to turn the battery
door fasteners, and then take of the battery case.

Replace with two new batteries.

Reinstall the battery case and tighten screws.

3.3. Replacing the Fuses

Warning

Summary of Contents for AX-C705

Page 1: ...oduct to see whether it is damaged in the shipment or not Check the materials to see whether they are the same as shown in the packing list Keep the packing materials for late delivery Standard and ch...

Page 2: ...l understanding of this Manual before use Make sure the requirements in the Manual are strictly followed and keep the Manual for reference whenever needs Mis operation may cause accidents and damages...

Page 3: ...the level of protection for the pulse compression typical installation location includes Three phase utility power supply equipment for any outdoor power lines or equipment Any outdoor transmission li...

Page 4: ...n stable 2 3 DC voltage measurement Turn the rotary switch to Insert black probe into COM terminal and insert red probe into _V terminal Connect the probes to the circuit under test and then read the...

Page 5: ...e electricity fully from all capacitor before testing continuity Turn the rotary switch to D press green button and select test Insert black probe into COM terminal and insert red probe into _V termin...

Page 6: ...d 250 HART Insert black probe into mA terminal and insert red probe into LOOP terminal Connect the probes to the circuit under test and then read the measured value when stable Press mA button to disp...

Page 7: ...A output Otherwise low impedance from other position of rotary switch will appear and cause current higher than 35mA to go through the loop circuit Turn the rotary switch to the screen displays OUTPUT...

Page 8: ...and latch the battery door before using the meter Note The new and old Batteries can not be mixed Max Voltage applied to output terminals about 32V Max current applied to output terminals about 25mA M...

Page 9: ...to IEC61326 1 Group 1 Class B Surge protection 8kV according to IEC61010 1 2001 Identification Mark CE Quality Standard Develops designs and manufactures in conformity with ISO 9001 4 2 General featu...

Page 10: ...ark DC voltage 50V 10mV 0 5 4 Measuring impedance 10 M standard value Common mode restrain 50Hz or 60Hz 100Db String mode restrain 50Hz or 60Hz 45dB Overload protection 600V peak peak value AC voltage...

Page 11: ...transmitter 20mA 0 000 22 000mA 0 001mA 0 2 4 external power supply 5 28V Max load 1k at 20mA LOOP Loop circuit power 24V 10 Max output current 25mA Max Voltage applied to output terminals about 32V...

Reviews: