3.  OPERATING PRINCIPLE 

 

The flow measuring principle is based on ultrasonic measurement method. The ultrasonic signal along the 

measuring section moves many times upstream and downstream between the ultrasonic sensors, who alternately 
performs transmitter and receiver functions. From the resulting time difference the flow rate is calculated. Measured 
flow rate is converted into the pulses quantity that is transferred in output pulse terminal. 

 

4.  MARKING AND SEALING 

 

4.1. Marking: 

 

 Marking of electronic unit 

 

There  are  following  information  on  the  front  panel  of  electronic  unit  - 

manufacturer’s  trade  mark,  type  of 

sensor,  serial  number,  year  of  manufacture,  EC-type  examination  certificate  number,  accuracy  class, 
environmental class by LST EN1434-1, electromagnetic and mechanical environmental class, enclosure protection 
class,  the  limiting  values  of  the  flow  rate  (qi,qp,qs),  output  pulse  value,  maximum  temperature  range  for  flow 
sensor, the maximum admissible working pressure. 
Numbers of terminal pins are marked close to the terminal 
 

Marking of flow sensor 

 

There are following information on the flow sensor: 

 

 -nominal diameter, 

            - arrow for indication of a flow direction  
 

  

4.2.

 Security seals

The following sealing of electronic unit is provided

 

Manufacturer‘s  adhesive  seal-sticker  on  the  access  to  the  adjustment  activation  jumper  (see  Annex  C, 

fig.C1, pos.1) and on the fixer on the cover protecting electronic wiring plate  (see Annex C, fig.C1, pos.2) 

 

The following measuring section sealing is provided

 

Manufacturer‘s  adhesive  seal-sticker  on  the  screws  of  protective  cover  of  flow  sensor  (see  Annex  C, 
fig.C2a; b; c; d; e).  
Manufacturer‘s  hanged  seals  on  ultrasonic  transducers  for  flow  sensors  with  steel  body  (see  Annex  C, 
fig.C2 f). 
 

Mounting seal 

 

After installation the case and cover of the electronic unit (see Annex C, fig.C1, pos.3) are sealed with two 

hanged seals  of heat supplier.

 

 

5.  SAFETY REQUIREMENTS 

 

Before beginning installation works you must to read this document and follow its instructions. 
 

 

Risk factors during the flowmeter installation and service is a heat conveying fluid flowing within flow sensor 

with inner pressure up to 1,6 (2,5) MPa and temperature up to 130 

0

C.  

 

To eliminate this risk the service and installation works must be performed by qualified technical personnel. 

Personnel must be familiar with appropriate technical documentation and general safety instructions. 

 

It is necessary to follow general safety requirements during installation and maintenance process. 

 

Protective grounding is not required, because housing is made from plastics, and conductive parts are not 
exposed to the surface. 

 

Safety guarantees at installation and service of meter is: 

 

- Reliable insulation of electrical circuits, 

 

- Hermetic fitting of primary flow sensor into the pipeline, 

 

- Reliable fastening of sub-assemblies of heat meter at installation. 

Warning!    Mounting  of  the  flowmeter  is  permissible  only  after  ensuring  of  absence  of  heat 

conveying fluid in the pipeline. 
              By  connecting  the  flow  sensor  to  the  external pulse  registration  device  which  is powered  via  AC 
230V check the proper grounding of external device. 

 
 
 

Summary of Contents for QALCOSONIC F2

Page 1: ...Axioma LEZ UAB ULTRASONIC FLOW SENSOR QALCOSONIC F2 TECHNICAL DESCRIPTION INSTALLATION AND USER INSTRUCTIONS PEF2V02 KAUNAS...

Page 2: ...classified as waste electronic equipment in terms of the European Directive 2012 19 EU WEEE Marking of electrical and electronic equipment in accordance with Article 14 2 of Directive 2012 19 EU This...

Page 3: ...aution If this equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired 1 APPLICATION FIELD Ultrasonic flow sensor QALCOSONIC F2 is desi...

Page 4: ...0 6 190 DN20 G 1 0 190 DN20 H 1 5 190 DN20 K 2 5 190 DN20 L 3 5 260 DN32 O 6 0 260 DN32 P 25 300 DN65 R 40 350 DN80 S 40 300 DN80 U 60 350 DN100 T 60 360 DN100 V Ratio of flow rate limits qp qi Code 1...

Page 5: ...9 8 G1 2 5 5 0 0 025 0 005 190 9 4 G1 DN20 3 5 7 0 0 035 0 017 260 4 G1 1 4 DN25 DN32 6 0 12 0 0 024 0 012 260 10 G1 1 4 DN25 DN32 6 0 12 0 0 06 0 012 260 10 G1 1 4 DN25 DN32 10 0 20 0 0 04 0 02 300 1...

Page 6: ...al power supply 12 V 42 V DC or 12 V 36 V 50 60 Hz AC 10 mA max internal backup battery size AA 3 6 V 2 4 Ah lithium battery Li SOCl2 service life not less than 12 years without data reading via a dig...

Page 7: ...The following measuring section sealing is provided Manufacturer s adhesive seal sticker on the screws of protective cover of flow sensor see Annex C fig C2a b c d e Manufacturer s hanged seals on ult...

Page 8: ...pulse cable is already installed Customer pulse cable Connection of the customer pulse cable It is recommended to use copper wire cable with the cross section 0 2 0 5 mm2 By means of tweezers remove...

Page 9: ...on in normal working mode pipeline must be under pressure and fully filled with heat conveying liquid The location and position of the flow sensor must be selected in such a way as to reduce the risk...

Page 10: ...in Figure 7 2 Fig 7 2 Test mode activation For working in this mode the calculator cannot be closed When the jumper J is set output pulses are formed Ending of verification mode Remove jumper J to le...

Page 11: ...PEF2V02 2017 11 20 11 Annex A Electrical wiring diagrams of flow sensor QALCOSONIC F2 Fig A1 Electrical wiring diagram...

Page 12: ...to the external power supply Table A1 Numbering of terminals Terminal No Destination 18 Output pulses of volume 19 GND Numbering of external power supply module terminals Terminal No Destination 54 24...

Page 13: ...sensor Fig B2 Sizes and dimensions of the flow sensor Fig B2 1 Threaded end connections G3 4 Fig B2 2 Threaded end connections G1 Mounting length L 110 mm or L 165 mm mounting length L 130 mm a b Fig...

Page 14: ...PEF2V02 2017 11 20 14 Annex B a Threaded end connections G1 1 4 b Flanged end connection DN25...

Page 15: ...PEF2V02 2017 11 20 15 Annex B c Flanged end connection DN32 Fig B2 4 Threaded end connections G1 1 4 a Flanged end connection DN25 b Flanged end connection DN32 c mounting length L 260 mm a b...

Page 16: ...PEF2V02 2017 11 20 16 c Fig B2 5 Threaded end connections G2 a flanged end connection DN40 b c two design options Mounting length L 300 mm...

Page 17: ...Brass housing DN65 PN16 DN65 PN25 DN80 DN100 L 300 300 350 350 D 185 185 200 220 H 185 185 200 220 D1 145 145 160 180 B 200 200 215 235 D2 18 18 18 18 n 4 8 8 8 b Dimensions of flow sensors DN65 DN80...

Page 18: ...e sticker 2 manufacturer warranty seal adhesive sticker 3 mounting seal a Flow sensor G 3 4 b Flow sensor G 1 or DN20 1 1 1 1 c Flow sensor G 1 DN25 or DN32 d Flow sensor G 2 or DN40 1 1 1 1 e Flow se...

Page 19: ...ed in the paragraph 2 of this document if transportation storage and operation conditions will be followed Warranty period 12 months from bringing into operation but not more than 18 months from manuf...

Reviews: