1kHz
Input sensitivity: 1.0V
Overload protection: 250 V DC/AC peak value.
15. Temperature Test
1.
Turn the function/range selection knob to Temperature
range.
2.
Insert the two ends of the temperature sensor into the “V
Ω
Hz” and “ uAmA” input terminal,
the positive end into the “ V
Ω
Hz” input terminal.
3.
Connect the sensor of the temperature cable to the surface or inside of the tested object,
displayed as the right picture.
4.
read the present test result from LCD
5.
Press “SELECT” button to choose F temperature measurement mode, and press “SELECT” button
again to choose C temperature measurement mode, circulated like this.
Note:
a.
Without the signal input, LCD automatically displays the normal temperature. Do not input any
other signal, caution to avoid damaging the instrument or endangering the safety of the users.
b.
Do not replace the temperature sensor at will. Otherwise the accuracy can’t be assured.
c.
Do not input the voltage at Temperature Range! Otherwise there is the danger to damage the
instrument.
Zakres
Dokładność
Rozdzielczość
-20
°
C - 400
°
C
±(1.0% + 5)
1
°
C
400
°
C - 1000
°
C
±(1.5% + 15)
0
°
F - 750
°
F
±(1.0% + 5)
1
°
F
750
°
F do 1832
°
F
±(1.5% + 5)
Overload protection: 0.5A/250V