The meter is a stable multimeter with 40mm LCD display, driven by battery. It's widely used on 
measuring DCV, ACV, DCA, ACA, resistance, capacitance, diode, transistor, continuity test 
,temperature auto power off /on and LCD back – light . It’s an ideal tool for lab, factory and family. 

2. SAFETY NOTE 

The meter meets the standards of IEC1010. Read the operation manual carefully before operation. 

1

Do not input limit over-ranged. 

2

The voltage below 36V is safety. To avoid electric shock, check whether the test leads are 

connected correctly, whether the insulation is good when measuring over 36DCV or 25ACV. 

3

Remove the test leads when changing function and range. 

4

To select correct function and range, beware of error operation.

 

5

Do not operate the meter if battery case and back cover is not fixed. 

6

Do not input voltage when measuring resistance. 

7

Remove test leads from test point and turn off the power before replacing battery and fuse. 

8

SAFETY SYMBOLS 

”EXISTS DANGEROUS VOLTAGE

”  GND

”DUAL INSULATION 

”THE OPERATOR MUST REFER TO THE MANUAL 

”LOW BATTERY 

3. CHARACTERISTIC 

3.1

GENERAL 

1-1

Display :LCD displaying. 

1-2

Max. displaying: 1999

3 1/2digit

auto polarity indication. 

1-3

Measuring method: dual slope A/D conversion. 

Summary of Contents for AX-572

Page 1: ...Digital Multimeter AX 572 Instruction manual...

Page 2: ...eads are connected correctly whether the insulation is good when measuring over 36DCV or 25ACV 3 Remove the test leads when changing function and range 4 To select correct function and range beware of...

Page 3: ...H 80 1 8 Power 9V 1 NEDA1604 6F22 or equivalent model 1 9 Size 175 93 55mm 1 10 Weight approx 400g including battery 3 2 TECHNICAL CHARACTERISTIC 2 1 Accuracy a rdg d at 23 5 R H 75 one year guarantee...

Page 4: ...nges Overload protection 250V DV or AC peak value at 200mV range 1000V DC or AC peak value at other ranges 2 2 2 ACV RANGE ACCURACY RESOLUTION 2V 1mV 20V 10mV 200V 0 8 5 100mV 750V 1 2 10 1V Input res...

Page 5: ...n 10 seconds Overload protection 0 2A 250V 20A 250V fast blown fuse 2 2 4 ACA RANGE ACCURACY RESOLUTION 2mA 1 0 15 1uA 200mA 2 0 5 100uA 20A 3 0 10 10mA Max measuring volt drop 200mV Max input current...

Page 6: ...0V DC and AC peak value NOTE at 200 range the test leads should be short circuit and measure the down lead resistance then subtract from the real measuring WARNING DO NOT input any voltage at resistan...

Page 7: ...1mA negative voltage is approx 3V Buzzer sounds the resistance is less than 70 20 open voltage is approx 3V Overload protection 250V DC or AC peak value Warning DO NOT input any voltage at this range...

Page 8: ...nge knob 5 20A current test jack 6 pole jack for capacitance temp and test accessory and less than 200mA current test jack 7 pole jack for capacitance temp test accessory and GND 8 pole jack for volt...

Page 9: ...ltage is unsure beforehand should set the range knob to the highest range then switch to a proper range according to the displayed value 2 If LCD displays 1 it means over range should set the range kn...

Page 10: ...the black test lead to COM jack and the red one to V jack 2 Set the range knob to a proper resistance range connect the test leads across to the resistance under measured NOTE 1 If the resistance val...

Page 11: ...e black test lead to COM terminal and the red one to V jack Note the polarity of red test lead is 2 Set the range knob to range connect the test leads to the diode under measured reading is the approx...

Page 12: ...d soft solvent to clean the meter do not use abrasive and alcohol 4 If do not operate it for a long time should take out the battery 4 1 When LCD displays symbol should replace the battery as below 4...