background image

What is in the case

• 

Axicon 12600, 12700, 12800 or 12900 2D verifier.

• 

Power supply and fly lead.

•  Mains lead.
• 

Wallet containing this documentation booklet, software CD and the calibration 

card.

Optional extras

Axicon also offer the following optional extras to be used in conjunction with the 

system:
• 

Installation qualification and operational qualification procedural 

documentation. 

• 

Performance Qualification Test Kit – Comprising a set of test cards traceable to 
international standards and full procedural instructions.

• 

Set of Reference Test Cards – traceable to international standards.

www.axicon.com

Page 5 of 24

Axicon Auto ID Limited | 12600 - 12900 user guide | June 2017

Summary of Contents for 12000

Page 1: ...ARCODE EXPERTS Church Road Weston on the Green Oxfordshire OX25 3QP UK Tel 44 0 1869 351155 Email sales axicon com Web www axicon com user guide Axicon 12600 12900 range 2D barcode verifiers Industry Partner ...

Page 2: ...www axicon com Page 2 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 3: ...the verifier 12 Calibration 13 Verification results 14 Using the 12800 12900 16 Software licence agreement 19 Annual servicing 20 Axicon Warranty 21 Certificate of conformance 23 Contact us 24 www axicon com Page 3 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 4: ...rated annually The unit should be returned to Axicon or to your approved Axicon reseller for this servicing The following table shows the various models in the Axicon 12000 series which can be used in conjunction with this software and their typical applications www axicon com Model Field of view Application 12000 24 x 18mm High resolution 2D codes 12500 40 x 24mm High to medium resolution 2D code...

Page 5: ...wing optional extras to be used in conjunction with the system Installation qualification and operational qualification procedural documentation Performance Qualification Test Kit Comprising a set of test cards traceable to international standards and full procedural instructions Set of Reference Test Cards traceable to international standards www axicon com Page 5 of 24 Axicon Auto ID Limited 126...

Page 6: ... process will check whether the NET framework is present and install it if necessary Insert the software CD into an appropriate drive and the Install program should start automatically If not run Setup exe Set the location where you would like the software installed and specify whether the installation should be for a single user or all users You will be asked to confirm the installation before it...

Page 7: ...vers that are needed will be installed automatically Running the software for the first time When running the software for the first time you will be asked to specify certain application options The application mode UDI HIBC relates to requirements of the Health Industry Barcode Standard so choose this if necessary Most users choose the Don t print report after each scan option You can always chan...

Page 8: ...es The software can be configured to meet your requirements when verifying both types of barcode Configuration of the 1D mode Double click on the 1D icon to display the configuration screen Software 1D www axicon com Page 8 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 9: ...an Details information relating to each of the scans will be provided individually To include an image of the barcode that has been verified select Image of Symbol The Report and Print Options section allows you to choose in which format the verification reports should be produced and where they will be saved automatically Select Metric Units at the right of this section if you would like measurem...

Page 10: ...Configuration of the 2D mode Double click on the 2D icon to display the configuration screen Software 2D www axicon com Page 10 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 11: ...and Reporting Options section you can choose how much information will be included in the verification report GS1 users should ensure that ISO 15415 option is selected The other options to the right of this are only usable with the Axicon 12800 and 12900 verifier which is designed to verify direct part marked symbols Saving and printing The saving of reports takes place automatically and you will ...

Page 12: ...t button on the verifier a live image will appear in the window Position the verifier so that the barcode is in its centre The Go Live button will now say Verify code Click on this or press the Start button again and the verification will begin After a few seconds the results will appear in the window with the ISO grade appearing in the centre Using the verifier www axicon com Page 12 of 24 Axicon...

Page 13: ...code are vertical in the camera window Click on the Start button of the verifier to activate the camera Go to Options then select Calibrate Remote The calibration dialogue box will be displayed Select the linear barcode and make sure the Rmin and Rmax values match those on your calibration card Set the calibration reminder to 15 days and then click on the calibrate button The calibration will take...

Page 14: ...r tabs provide the results set out in slightly different ways The most interesting tab is called SRP for Scan Reflectance Profile and this provides a graph showing the brightness of the symbol measured across its width The high points represent the spaces while the lowest points represent the bars The round green arrow buttons allow you to move through the ten different scans and as you do this th...

Page 15: ...alue for each element and the letter grades and percentage values for Modulation Defects and Decodability Verifying 2D symbols You follow the same procedure as for linear symbols after making sure that you have selected Enable 2D below the blue 2D configuration icon at the top of the command window The results screen will now include an image of the code with additional details that are required b...

Page 16: ...arked on more unusually shaped objects or where the marking is slightly recessed The appropriate buttons to change the angle of illumination appear at the top of the main command window to the right of the green x icon You should first try to see if you can obtain an acceptable pass grade by using the ISO IEC 15415 standard and then if not by using the ISO IEC TR 29158 AIM DPM approach The 2D conf...

Page 17: ...both the GS1 ISO IEC 15415 and AIM DPM approaches are selected the verification report will include both sets of results www axicon com Page 17 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 18: ...two red dots overlap The red laser spot will move up or down the image as the handwheel is turned Best focus will be achieved when the red laser spot is aligned with the small red ring that is shown in the middle of the image and when these two coincide the red ring will change to a green coloured ring and the Verify code button can then be clicked to decode and verify the symbol Caution Take care...

Page 19: ...www axicon com Page 19 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 20: ...n it you with the latest version of the software on CD together with a new calibration card and a new user guide that will include the new certificate of ISO IEC conformance You will then need to calibrate the verifier before you use it again The latest software can always be downloaded free of charge from our website www axicon com To arrange for your verifier to be serviced 1 Contact us on Telep...

Page 21: ...the equipment is returned without the original transit case we will automatically charge an additional sum for the replacement of this packaging We reserve the right to charge an inspection fee for any equipment returned under warranty for which no fault is found Please contact us to obtain an RMA return to manufacturer authorisation number before returning any equipment This number must be quoted...

Page 22: ...www axicon com Page 22 of 24 Axicon Auto ID Limited 12600 12900 user guide June 2017 ...

Page 23: ...quipment when used with the latest verifier software and configured appropriately has the features to meet the requirements of the United States Food and Drug Administration regulation Title 21 CFR Part 11 This verifier has been quality assurance checked today and conforms to these ISO IEC standards This declaration of conformity is valid for a year Issue date Expiry date Authorised signatory Sign...

Page 24: ...0 1869 351 155 Fax 44 0 1869 351 205 Email info axicon com Website www axicon com As Axicon continually improves its products their specifications and features are subject to change without notice All trademarks are acknowledged as belonging to their respective companies Your Axicon distributor ...

Reviews: