![AWS MSC-200 User Manual Download Page 7](http://html1.mh-extra.com/html/aws/msc-200/msc-200_user-manual_3035774007.webp)
free of any air currents or vibration
•
Problem:
Displayed weight is inaccurate
•
Solution:
Perform a calibration adjustment
XII. Error Codes
1. OUTZ
- Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. 88888
- Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. L
- Batteries are low. Replace the batteries.
MSC-200
MSC-650
Max. Capacity
200g
650g
Readability
0.01g
0.1g
Min. Capacity
0.03g
0.2g
Power
USB Cable / Internal Rechargeable
Platform Dimensions
2.3 x 2.8”
Scale Dimensions
2.3 x 4.8 x 1.5”
Warranty
10 Year Limited Warranty
XIII. Capacity/Readability Charts
MSC-200
MSC-650
200x0.01g
7.05x0.001oz
1000x0.05ct
3089x0.1gn
650x0.1g
22.92x0.01oz
20.89x0.01ozt
417.9x0.1dwt