AWS MSC-200 User Manual Download Page 7

 free of any air currents or vibration

• 

 

Problem:

 Displayed weight is inaccurate

• 

 

Solution:

 Perform a calibration adjustment

XII.  Error Codes

1.  OUTZ

 - Zero range has shifted. Perform calibration 

adjustment. If the problem persists, the load cell may 

be permanently damaged due to overloading.

2.  88888

 - Maximum capacity exceeded. Remove the 

extra weight to avoid damaging the load cell.

3.  L 

- Batteries are low. Replace the batteries.

MSC-200

MSC-650

Max. Capacity

200g

650g

Readability

0.01g

0.1g

Min. Capacity

0.03g

0.2g

Power

USB Cable / Internal Rechargeable

Platform Dimensions

2.3 x 2.8”

Scale Dimensions

2.3 x 4.8  x 1.5”

Warranty

10 Year Limited Warranty

XIII.  Capacity/Readability Charts

MSC-200

MSC-650

200x0.01g

7.05x0.001oz

1000x0.05ct

3089x0.1gn

650x0.1g

22.92x0.01oz

20.89x0.01ozt

417.9x0.1dwt