3. 

You may now fill the container with the quantity to 

be weighed. The display will show the net weight of 

the contents.

4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 

press the

 T

 key again to re-zero the scale.

VII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart below), adjustment is required.

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

MCD-100

0.02g

100g

OIML M2

MCD-500

0.2g

500g

OIML M2

VIII.  Calibration Adjustment

1.  Press and hold the 

M

 key until the display shows 

“CAL”.

2.  Press the 

M

 key again. The display will flash “CAL” 

followed by the calibration weight value (see chart).

3.  Place the calibration weight on the center of the 

weighing platform. Make sure you are using the 

cor

rect weight for your specific model.

Summary of Contents for MiniCD MCD-100

Page 1: ...American Weigh Scales MCD 100 100 x 0 01g MCD 500 500 x 0 1g User Manual MiniCD Copyright 2010 American Weigh Scales Inc All rights reserved Rev 2 0 ...

Page 2: ...rming calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or malfunction Avoid using your scale in close proximity to strong radio wave emitting devices such as computers cash registers and cordless phones Always weigh on a flat and level surface free from vibrations and drafts The corner...

Page 3: ...ackaging that may be in place to prevent battery drainage during shipping storage 2 When installing the batteries make sure the positive and negative contacts are properly aligned Install Negative Side Up Install Positive Side Up Postive Terminal CR2032 Button Cell Battery Installation I n s t a l l N e g a t i v e S i d e U p I n s t a l l P o s i t i v e S i d e U p C R 2 0 3 2 B u t t o n C e l...

Page 4: ... to change the scale s displayed unit of measure 2 If the scale is turned off then back on it will default to the last unit used MCD 100 Units g gram oz ounce ct carat gn grain MCD 500 Units g gram oz ounce ozt troy ounce dwt pennyweight VI Using the Tare Feature You can use this scale s tare feature to subtract the weight of an empty container or vessel for net weight determination NOTE The weigh...

Page 5: ...moving the scale from one location to another To check calibration 1 Place the required calibration weight on the scale and note the displayed value 2 If the displayed value is not within acceptable tolerance see chart below adjustment is required Model Acceptable Tolerance Calibration Weight Recommended Accuracy Class MCD 100 0 02g 100g OIML M2 MCD 500 0 2g 500g OIML M2 VIII Calibration Adjustmen...

Page 6: ...eight is inaccurate Solution Perform a calibration adjustment X Error Codes 1 OUTZ Zero range has shifted Perform calibration adjustment If the problem persists the load cell may be permanently damaged due to overloading 2 88888 Maximum capacity exceeded Remove the extra weight to avoid damaging the load cell 3 L Batteries are low Replace the batteries MCD 100 MCD 500 Max Capacity 100g 500g Readab...

Reviews: