AWS CQ-350 User Manual Download Page 5

VI. CALIBRATION

Calibration is NOT required before first use. Only

calibrate if the scale is not weighing properly.

The CQ-500 calibrates with a 500g weight*

The CQ-350 calibrates with a 350g weight*

*Weight not included.

1. Turn on the scale by pressing the [ 

ON/OFF

 ] key.

2. Wait until the LCD displays “0.0” , then press and 

HOLD the [ 

CALI

 ] key for 5 seconds. The LCD will 

display “0”. You may now release the [ 

CALI

 ] key. 

3. The display will flash the required calibration weight. 

Place the required weight on the platform. (see above)

4. The display will show “PASS” and return to normal 

weighing mode.

5. Put away the calibration weight and press

ON/OFF

 ] to turn off the scale. Calibration is finished!

VII. TROUBLESHOOTING

The primary reasons for inaccuracy or malfunction 

are low batteries, incorrect calibration, overload or 

operating on an unstable surface. Please keep this in 

mind and maintain and operate your scale properly. 

This scale is a precise instrument and must be 

handled with the utmost care and caution.

[Display]

  Elimination of errors

----------------------------------------------- 

[OUT2]  

  Recalibrate the scale (Section VI).

[EE] 

 

You are overloading the scale. Remove the  

 

excessive load immediately!

[LO]  

 

Replace the batteries.