AWS CHROME-1KG User Manual Download Page 5

4.  To return to normal weight display (gross weight), 

remove any items from the weighing platform and 

press the

 TARE

 key again to re-zero the scale.

VII.  Using the Counting Feature

1.  Press the 

ON/OFF

 key to turn the scale on.

2.  Place a sample quantity of the items being weighed 

on the platform. This quantity should either be 25, 

50, 75, or 100 pieces.

3.  Once the weight is stable, press and hold the 

PCS

 

key to enter counting mode setup. 

4.  The current sample size will be displayed on the 

LCD (ie. “P  25”).

5.  Press the 

UNIT

 key to change it between 25, 50, 

75, or 100 and press 

PCS

 to confirm. The count will 

be displayed on the LCD. You may now add more 

items. As items are added or removed the count will 

update automatically.

VIII.  Calibration

Calibration and adjustment are performed at the factory. 

However, the weighing range can shift slightly during 

shipping. Also, due to changes in local acceleration of gravity 

around the world, adjustment is recommended when moving 

the scale from one location to another. To check calibration:

1.  Place the required calibration weight on the scale 

and note the displayed value.

2.  If the displayed value is not within acceptable 

tolerance (see chart on the next page), adjustment 

is required.