AWS BCM-100 User Manual Download Page 6

BCM-Series Manual

Model

Acceptable 

Tolerance (±)

Calibration 

Weight

Recommended 

Accuracy Class

BCM-100

0.02g

100g

OIML M2

BCM-650

0.2g

500g

OIML M2

VIII.  Adjustment

1.  Press and hold the 

MODE

 key until the display 

shows “CAL”

2.  Make sure nothing is on the scale platform then 

press the 

MODE

 key again.

3. 

The display will flash “CAL” followed by the required 

calibration weight (see chart above).

4.  Place the required calibration weight on the platform. 

The display will flash “PASS” then return to normal 

weighing mode. Adjustment is complete!

IX. Troubleshooting

• 

 

Problem:

 Scale will not turn on

• 

 

Solution: 

Check the batteries and battery contacts.

• 

 

Problem: 

Displayed weight fluctuates randomly

• 

 

Solution: 

Make sure your workspace is stable and    

 free of any air currents or vibration. In some cases, 

nearby radio wave emitting devices can cause 

fluctuations.

• 

 

Problem:

 Displayed weight is inaccurate

• 

 

Solution:

 Perform a calibration adjustment. If 

you are still having trouble, contact our warranty 

department. 

Summary of Contents for BCM-100

Page 1: ...American Weigh Scales BCM 100 100g x 0 01g BCM 650 650g x 0 1g User Manual BCM Series Copyright 2012 American Weigh Scales Inc All rights reserved Rev 1 2...

Page 2: ...or 30 60 seconds before performing calibration if available Store your scale in a clean dry location Dust dirt and moisture can accumulate on the weighing sensors and electronics causing inaccuracy or...

Page 3: ...scription 1 ON OFF Turns the scale on or off 2 Mode Key Tap to change the displayed unit of measure Hold to enter calibration see VII 3 Tare Key Re zeros the scale Used to remove the weight of a tray...

Page 4: ...properly aligned IV Operation 1 Place the scale on a flat level surface and press the ON OFF key to turn the scale on 2 Once the start up process completes and the scale goes to zero you are ready to...

Page 5: ...e weighed The display will show the net weight of the contents 4 To return to normal weight display gross weight remove any items from the weighing platform and press the TARE key again to re zero the...

Page 6: ...ace the required calibration weight on the platform The display will flash PASS then return to normal weighing mode Adjustment is complete IX Troubleshooting Problem Scale will not turn on Solution Ch...

Page 7: ...3 Lo Batteries are low Replace the batteries XI Specifications BCM 100 BCM 650 Max Capacity 100g 650g Readability 0 01g 0 1g Min Weight 0 03g 0 2g Power 2 x 1 5V AAA Alkaline Dimensions 2 7 x 4 0 x 0...

Page 8: ...BCM Series Manual American Weigh Scales Inc Norcross GA 30092 Made In China to our exact specifications and quality control...