AMW-BENCH
OFF
key to turn the scale on.
3. Press and hold the
MODE
key until the display
shows “CAL”.
4. Press the
MODE
key again. The display will flash
“CAL” followed by the calibration weight value (see
chart on page 5).
5. Place the calibration weight on the center of the
weighing platform. Make sure you are using the
correct weight for your specific model.
6. Once the weight is accepted, the display will show
“PASS” then return to normal operation. Turn the
scale off and return the switch on the bottom to
Auto-Off Enabled or Auto-Off Disabled.
X. Troubleshooting
•
Problem:
Scale will not turn on
•
Solution:
Check the batteries
•
Problem:
Displayed weight fluctuates randomly
•
Solution:
Make sure your workspace is stable and
free of any air currents or vibration
•
Problem:
Displayed weight is inaccurate
•
Solution:
Perform a calibration adjustment
XI. Error Codes
1. OUTZ
- Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. o-Ld
- Maximum capacity exceeded. Remove the